Intracardiac Electrograms and Sensing Test Signals: Electrophysiological, Physical, and Technical Considerations
- 1 November 1985
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 8 (6) , 870-888
- https://doi.org/10.1111/j.1540-8159.1985.tb05907.x
Abstract
Until now. the characterization of intracardiac electrograms was carried out mainly in the frequency domain by which filtering and sensitivity testing was decidingly influenced. The typical structure of the intracardiac electrogram derived with unipolar electrodes that is normally responsible for synchronization can be explained theoretically by assuming two dipoles perpendicular to each other moving along with nn excitation wave. An abstraction of this typical structure yields, as a sensitivity test signal, an asymmetric triangle with a 4-ms decay time and a 16-ms rise time. A correlation of different test signals and real heart signals applied to 16 different pacemakers demonstrates that the proposed test signal is superior to all others known so far.Keywords
This publication has 15 references indexed in Scilit:
- Optimal Pacemaker Sensing with Respect to Amplitude and Slew Rate of Intracardiac Electrograms: Theoretical Analysis by Computer SimulationPacing and Clinical Electrophysiology, 1984
- Long-term Comparison of Unipolar and Bipolar Pacing and Sensing, Using a New Multiprogrammable Pacemaker SystemPacing and Clinical Electrophysiology, 1983
- Sensing Characteristics of the Right Atrial Appendage ElectrodePacing and Clinical Electrophysiology, 1983
- Electrophysiological Properties of a New Permanent Endocardial Lead for Uni‐ and Bipolar PacingPacing and Clinical Electrophysiology, 1982
- Myopotential Inhibition of Unipolar QRS‐inhibited (VVI) Pacemakers, Assessed by Ambulatory Holter Monitoring of the ElectrocardiogramPacing and Clinical Electrophysiology, 1980
- Characteristics of Intracardiac Electrograms II: Atrial Endocardial ElectrogramsPacing and Clinical Electrophysiology, 1980
- Compendium of Pacemaker Technology. IIPacing and Clinical Electrophysiology, 1979
- The Interdependence Between Electrogram, Total Electrode Impedance and Pacemaker Input Impedance Necessary to Obtain Adequate Functioning of Demand PacemakersPacing and Clinical Electrophysiology, 1979
- PACE DATA CARD®Pacing and Clinical Electrophysiology, 1979