Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster
Open Access
- 5 October 2006
- journal article
- Published by Wiley in Acta Ophthalmologica Scandinavica
- Vol. 85 (1) , 84-87
- https://doi.org/10.1111/j.1600-0420.2006.00774.x
Abstract
Purpose: This study aimed to demonstrate how the level of accuracy in intraocular lens (IOL) power calculation can be improved with optical biometry using partial optical coherence interferometry (PCI) (Zeiss IOLMaster) and current anterior chamber depth (ACD) prediction algorithms. Methods: Intraocular lens power in 461 consecutive cataract operations was calculated using both PCI and ultrasound and the accuracy of the results of each technique were compared. To illustrate the importance of ACD prediction per se, predictions were calculated using both a recently published 5-variable method and the Haigis 2-variable method and the results compared. All calculations were optimized in retrospect to account for systematic errors, including IOL constants and other off-set errors. Results: The average absolute IOL prediction error (observed minus expected refraction) was 0.65 dioptres with ultrasound and 0.43 D with PCI using the 5-variable ACD prediction method (p < 0.00001). The number of predictions within ± 0.5 D, ± 1.0 D and ± 2.0 D of the expected outcome was 62.5%, 92.4% and 99.9% with PCI, compared with 45.5%, 77.3% and 98.4% with ultrasound, respectively (p < 0.00001). The 2-variable ACD method resulted in an average error in PCI predictions of 0.46 D, which was significantly higher than the error in the 5-variable method (p < 0.001). Conclusions: The accuracy of IOL power calculation can be significantly improved using calibrated axial length readings obtained with PCI and modern IOL power calculation formulas incorporating the latest generation ACD prediction algorithms.Keywords
This publication has 19 references indexed in Scilit:
- Prediction of the effective postoperative (intraocular lens) anterior chamber depthPublished by Wolters Kluwer Health ,2006
- Optimization of biometry for intraocular lens implantation using the Zeiss IOLMasterActa Ophthalmologica Scandinavica, 2005
- Optimization of biometry for intraocular lens implantation using the Zeiss IOLMasterActa Ophthalmologica Scandinavica, 2005
- Calibration of axial length measurements with the Zeiss IOLMasterJournal of Cataract & Refractive Surgery, 2005
- Ocular Component Measurement Using the Zeiss IOLMasterOptometry and Vision Science, 2004
- A new non-contact optical device for ocular biometryBritish Journal of Ophthalmology, 2002
- Immersion A-scan compared with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Biometry of cataractous eyes using partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™*Ophthalmic and Physiological Optics, 2001
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992