X-ray Beam Equalization: Feasibility and Performance of an Automated Prototype System in a Phantom and Swine
- 1 December 2001
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 221 (3) , 668-675
- https://doi.org/10.1148/radiol.2213010183
Abstract
To investigate the feasibility of an automated implementation of a beam equalization technique and to evaluate the experimental performance of the prototype system.X-ray beam equalization involved the process of low-dose image acquisition, attenuator thickness calculation, mask generation, mask positioning, equalized image acquisition, and mask reshaping. The entire equalization process was performed in approximately 7 seconds. The equalized images were assessed both qualitatively and quantitatively by using a humanoid phantom and a swine animal model. The general image quality was assessed for the ability to visualize arterial branches and other anatomic structures. The level of equalization was quantitatively assessed by segmenting the images into an 8 x 8 matrix of square regions.The ratio of the root-mean-squared variance for the equalized and unequalized images from humanoid phantom and swine animal studies was 0.49 and 0.59, respectively. Furthermore, qualitative assessment of the images showed substantial improvement in image quality and visualization of arterial branches after beam equalization in phantom and animal studies.Automated area beam equalization is feasible and improves image quality in previously underpenetrated regions of the image.Keywords
This publication has 15 references indexed in Scilit:
- Filter wheel equalization for chest radiography: A computer simulationMedical Physics, 1995
- Filter wheel equalization in DSA: Simulation resultsMedical Physics, 1993
- AMBER: a scanning multiple-beam equalization system for chest radiography.Radiology, 1988
- Geometrical properties of a digital beam attenuator systemMedical Physics, 1987
- Digital beam attenuator technique for compensated chest radiography.Radiology, 1986
- A scanning system for chest radiography with regional exposure control: Practical implementationMedical Physics, 1983
- A scanning system for chest radiography with regional exposure control: Theoretical considerationsMedical Physics, 1983
- Digitally Controlled Beam AttenuatorPublished by SPIE-Intl Soc Optical Eng ,1982
- A scanning equalization system for improved chest radiography.Radiology, 1982
- Primary X-Ray DodgingRadiology, 1971