Validation of a New Breathing Simulator Generating and Measuring Inhaled Aerosol with Adult Breathing Patterns
- 1 June 2000
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine
- Vol. 13 (2) , 139-146
- https://doi.org/10.1089/089426800418668
Abstract
The use of breathing simulators for the in vitro determination of the inhaled mass of drug from nebulizers has become widely accepted. Their use is, however, based on the assumption that there is a correlation between the in vitro and in vivo inhaled mass of drug. The aim of the study was therefore to investigate whether a new breathing simulator - the MIMIC Breathing Emulator (Medic-Aid Limited, Bognor Regis, UK) - could accurately emulate the in vivo inhaled mass of budesonide suspension for nebulization. Eight adult healthy subjects were included. Each subject inhaled for 2 min from a Spira Module 1 jet nebulizer (Respiratory Care Center, Hämeenlinna, Finland), charged with 1.0 mg of budesonide suspension for nebulization (0.5 mg mL-1, 2 mL suspension, AstraZeneca, Sweden) and supplied with an inhaled mass filter between the nebulizer and the subject. The breathing patterns were recorded during the nebulization and simulated in vitro at two different experimental sites (simulations A and B) with the breathing simulator. With the patients breathing through the filters (in vivo test), inhaled mass of budesonide averaged 103.6 μg. The two in vitro experiments using the simulator revealed similar results with in vitro simulation A equal to 101.0 μg and simulation B 99.1 μg. There were no statistically significant differences between the in vivo results and those of in vitro simulation A. Results were significantly different for simulation B (p = 0.032) although the difference was less than 4.5%. These data indicate that the breathing simulator can be used to accurately simulate sine waveforms, human breathing patterns, and the in vitro and in vivo inhaled mass of budesonide suspension for nebulization.Keywords
This publication has 13 references indexed in Scilit:
- Effects of Equipment Dead Space and Pediatric Breathing Patterns on Inhaled Mass of Nebulized BudesonideJournal of Aerosol Medicine, 1999
- The Conventional Ultrasonic Nebulizer Proved Inefficient in Nebulizing a SuspensionJournal of Aerosol Medicine, 1999
- Design, validation and initial testing of the Electronic LungTM deviceJournal of Aerosol Science, 1998
- In Vitro Determination of Inhaled Mass and Particle Distribution for Budesonide Nebulizing SuspensionJournal of Aerosol Medicine, 1998
- In Vitro Performance Testing of Three Small Volume-Holding Chambers under Conditions That Correspond with Use by Infants and Small ChildrenJournal of Aerosol Medicine, 1997
- Effect of Age Related Breathing Parameters on the Performance of a New Nebulizer System: An In-Vitro StudyJournal of Aerosol Medicine, 1995
- Drug Delivery by Nebulization: "Reality Testing"Journal of Aerosol Medicine, 1994
- Drug Delivery SystemsJournal of Aerosol Medicine, 1994
- Drug delivery from jet nebulisers.Archives of Disease in Childhood, 1992
- Drug Delivery via Aerosol Systems: Concept of "Aerosol Inhaled"Journal of Aerosol Medicine, 1991