Repeatability and Reproducibility of Central Corneal Thickness Measurement With Pentacam, Orbscan, and Ultrasound
Top Cited Papers
- 1 October 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Optometry and Vision Science
- Vol. 82 (10) , 892-899
- https://doi.org/10.1097/01.opx.0000180817.46312.0a
Abstract
The purpose of this study was to compare central corneal thickness (CCT) measurements obtained with a novel rotating Scheimpflug camera (Pentacam; Oculus) with scanning slit topography (Orbscan; Bausch & Lomb), and with ultrasound pachymetry (SP-2000; Tomey). CCT in 30 healthy eyes was measured twice with each modality by 2 independent observers in random order. The results from scanning slit topography are given both with and without multiplication with the "acoustic correction factor" of 0.92. In addition, the displayed images from the rotating Scheimpflug camera and scanning slit topography were used to calculate the signal difference-to-noise ratios (SD/N) between cornea and background signal. The mean CCT values as determined with the different modalities (± standard deviation) were: 542 ± 29 μm, 576 ± 37 μm, 530 ± 34 μm, and 552 ± 32 μm for rotating Scheimpflug imaging, for uncorrected and for corrected scanning slit pachymetry, and for ultrasound, respectively. The differences between modalities (± 95% limits of agreement) were –9.8 ± 31 μm between rotating Scheimpflug and ultrasound, 24 ± 31.2 μm between scanning slit and ultrasound, and 33 ± 27 μm between scanning slit and rotating Scheimpflug imaging. The limits of agreement for within and between observer effects were within 4.2% of the absolute CCT values for scanning slit and ultrasound and within 2.2% for the rotating Scheimpflug imaging. The rotating Scheimpflug camera showed similar SD/N ratios but steeper edges of the corneal surfaces in the intensity profile plots. In the assessment of normal corneas, the Pentacam measured CCT values closer to ultrasound pachymetry and with less variability compared with Orbscan. The (interobserver) reproducibility with the Pentacam was highest of all 3 modalities.Keywords
This publication has 53 references indexed in Scilit:
- Corneal thickness measurement by confocal microscopy, ultrasound, and scanning slit methodsAmerican Journal of Ophthalmology, 2004
- Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan systemOphthalmology, 2004
- Comparison of Orbscan and Ultrasound pachymetry in the measurement of central corneal thicknessEye, 2004
- Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusisBritish Journal of Ophthalmology, 2004
- Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thicknessBritish Journal of Ophthalmology, 2004
- A new non-contact optical device for ocular biometryBritish Journal of Ophthalmology, 2002
- The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™*Ophthalmic and Physiological Optics, 2001
- Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implantBritish Journal of Ophthalmology, 2000
- Clinical assessment of anterior chamber depthOphthalmic and Physiological Optics, 1998
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986