Clinical blood flow quantification with segmented k‐space magnetic resonance phase velocity mapping
Open Access
- 23 December 2002
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 17 (1) , 65-71
- https://doi.org/10.1002/jmri.10231
Abstract
Purpose: To evaluate the accuracy of segmented k‐space magnetic resonance phase velocity mapping (PVM) in quantifying aortic blood flow from through‐plane velocity measurements.Materials and Methods: Two segmented PVM schemes were evaluated, one with seven lines per segment (seg‐7) and one with nine lines per segment (seg‐9), in twenty patients with cardiovascular disease. A non‐segmented (non‐seg) PVM acquisition was also performed to provide the reference data.Results: There was agreement between the aortic flow curves acquired with segmented and non‐segmented PVM. The calculated systolic and total flow volume per cycle from the seg‐7 and the seg‐9 scans correlated and agreed with the flow volumes from the non‐seg scans (differences < 5%). Sign tests showed that there were no statistically significant differences (P‐values < 0.05) between the segmented and the non‐segmented PVM measurements. Seg‐9, which was the fastest among the three sequences, provided adequate spatial and temporal resolution (> 10 phases per cycle).Conclusion: Segmented k‐space PVM shows great clinical potential in blood flow quantification. J. Magn. Reson. Imaging 2003;17:65–71.Keywords
This publication has 32 references indexed in Scilit:
- Evaluation of the Precision of Magnetic Resonance Phase Velocity Mapping for Blood Flow MeasurementsJournal of Cardiovascular Magnetic Resonance, 2001
- Quantification of mitral regurgitation with MR phase-velocity mapping using a control volume method.Journal of Magnetic Resonance Imaging, 1998
- Slice location dependence of aortic regurgitation measurements with MR phase velocity mappingMagnetic Resonance in Medicine, 1997
- Severity of aortic regurgitation: interstudy reproducibility of measurements with velocity-encoded cine MR imaging.Radiology, 1992
- Vena caval flow: assessment with cine MR velocity mapping.Radiology, 1990
- Quantitative flow measurements on phantoms and on blood vessels with MRMagnetic Resonance in Medicine, 1988
- In-plane flow velocity quantification along the phase encoding axis in MRIMagnetic Resonance Imaging, 1988
- Quantitative phase-flow MR imaging in dogs by using standard sequences: comparison with in vivo flow-meter measurementsAmerican Journal of Roentgenology, 1987
- Measurement of Flow with NMR Imaging Using a Gradient Pulse and Phase Difference TechniqueJournal of Computer Assisted Tomography, 1984
- A flow velocity zeugmatographic interlace for NMR imaging in humansMagnetic Resonance Imaging, 1982