Replacement correction factors for photon and electron dose measurements
- 1 November 1985
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 12 (6) , 785-787
- https://doi.org/10.1118/1.595665
Abstract
A "gradient correction factor" and an "electron fluence correction factor" are used in the AAPM Task Group 21 dosimetry protocol, the production of the two being the "replacement correction factor". The separation of two physical processes, implied by the concepts, is theoretically unfounded. The proof of this statement rests on Fano''s theorem, the conditions for which constitute a case without variations in fluence or its directional and energy distributions. Deviations from these conditions represent gradients in the radiation field, affect the response of a measurement cavity, and do so by perturbing the fluence of electrons that deposit energy in the cavity. Thus, in beams of both electrons and photons, corrections are required when fluence variations exist in the vicinity of the cavity. The rationale for an electron fluence correction based on "in-scattering" and "obliquity" is weak, since the effects only occur when gradients are present. The choice of values for the two correction factors at the depth of maximum dose are particularly inconsistent with the actual characteristics of the radiation field.Keywords
This publication has 2 references indexed in Scilit:
- Dose optimization with computer-controlled gantry rotation, collimator motion and dose-rate variationInternational Journal of Radiation Oncology*Biology*Physics, 1983
- Rotation therapy using a novel high-gradient filter.Radiology, 1982