Depiction of Low-Contrast Detail in Digital Radiography
- 1 July 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 41 (7) , 593-599
- https://doi.org/10.1097/01.rli.0000223765.73182.55
Abstract
We sought to evaluate the low-contrast performance of a newly developed needle image plate/line scanner (NIP) computed radiography system in comparison with a standard powder image plate/flying-spot scanner (PIP) system. A total of 36 images of a CDRAD phantom, simulating low-contrast structures with different drill holes of different diameters, were obtained with both imaging systems using 9 different exposure variables. All images had window and level set to generate consistent density and contrast. In addition, multiscale contrast-dependent contrast amplification was applied to some of the images. All images obtained were printed and presented to a total of 10 observers (5 radiologists, 5 engineers/physicists), who were blinded to both the image plate and parameter setting used. The smallest detectable drill hole depth (= contrast) correctly identified was recorded for each diameter. The median values observed were calculated and tested for statistical differences between PIP and NIP using Student t test for matched pairs (level of significance P < or = 0.05). At all but 2 settings of the variables, NIP images depicted significantly lower contrast levels (= lower depth of drill holes) compared with PIP images. The 2 settings also showed a trend towards better low contrast depiction with NIP. In no case was low contrast performance better using PIP images. Images obtained with needle image plate/line scanner provide superior low contrast performance compared with the images obtained with powder image plate/flying-spot scanner.Keywords
This publication has 26 references indexed in Scilit:
- Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practiceEuropean Radiology, 2005
- Computed Radiography and Direct RadiographyInvestigative Radiology, 2005
- Reflective and transmissive CR ScanHead technology on needle image platesPublished by SPIE-Intl Soc Optical Eng ,2005
- A new fast and large area neutron detector using a novel image plate readout techniqueApplied Radiation and Isotopes, 2004
- Measurement of the detective quantum efficiency (DQE) of digital x-ray imaging devices according to the standard IEC 62220-1Published by SPIE-Intl Soc Optical Eng ,2004
- New high-speed scanning technique for computed radiographyPublished by SPIE-Intl Soc Optical Eng ,2002
- A new needle-crystalline computed radiography detectorJournal of Digital Imaging, 2000
- Digital image processingEuropean Radiology, 1997
- Comparison of receiver operating characteristic and forced choice observer performance measurement methodsMedical Physics, 1995
- Multiscale image contrast amplification (MUSICA)Published by SPIE-Intl Soc Optical Eng ,1994