Monte Carlo simulations were performed to determine whether the multicomponent T2 distribution of tissue can be estimated accurately from T2 decay data acquired in vivo. Simulated data were generated for white matter, fast twitch muscle, and breast tissue. The signal‐to‐noise ratio, number of data samples, and minimum echo time were varied from the experimental conditions currently achievable with MRI to those achievable for in vitro experiments. Data were fitted by a distribution of T2 values using the T2NNLS algorithm, and statistics characterizing the estimated T2 components were determined. Current MRI techniques were found to provide conditions insufficient for accurate multicomponent T2 analysis on a pixel‐by‐pixel basis. However, volume localization methods that measure T2 decay from a large volume of interest have potential for this analysis. These results illustrate a general framework for development of new techniques to measure T2 decay accurately in vivo.Keywords
Funding Information
- National Cancer Institute of Canada
- General Electric Medical Systems, Canada
This publication has 19 references indexed in Scilit:
- In vivo visualization of myelin water in brain by magnetic resonanceMagnetic Resonance in Medicine, 1994
- Practical T2 quantitation for clinical applicationsJournal of Magnetic Resonance Imaging, 1992
- Quantitative Magnetic Resonance Imaging Parameters and Their Relationship to Mammographic PatternJNCI Journal of the National Cancer Institute, 1992
- Lipid‐suppressed single‐and multisection proton spectroscopic imaging of the human brainJournal of Magnetic Resonance Imaging, 1992
- Investigation of analysis techniques for complicated NMR relaxation dataJournal of Magnetic Resonance (1969), 1991
- Pulsed NMR relaxometry of striated muscle fibersMagnetic Resonance in Medicine, 1991
- Quantitative interpretation of NMR relaxation dataJournal of Magnetic Resonance (1969), 1989
- Information available and unavailable from multiexponential relaxation dataJournal of Magnetic Resonance (1969), 1989
- Analysis of biological NMR relaxation data with continuous distributions of relaxation timesJournal of Magnetic Resonance (1969), 1986
- CONTIN: A general purpose constrained regularization program for inverting noisy linear algebraic and integral equationsComputer Physics Communications, 1982