Polyethylene wear with an entirely HA-coated total hip replacement 79 hips followed for 5 years
Open Access
- 1 January 1998
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 69 (3) , 253-258
- https://doi.org/10.3109/17453679809000925
Abstract
We measured the eccentricity of the femoral head in the metal backing annually during 5 years in 79 consecutive total hip replacements (73 patients). The mean age of the patients was 57 (32–73) years and the female/male ratio 63/22. The prostheses were entirely coated with hydroxyapatite. Modular 32 mm stainless steel heads and hemispherical, self-tapping screw cups with polyethylene liners were used. We found accelerating eccentricity throughout the observation period. The mean eccentricity at 5 years was 0.71 mm (95% CI 0.53–0.90), resulting in a mean eccentricity rate of 0.14 mm (0.11 -0.18) per year. The 5-year eccentricity was 0.5 mm or less in 45 hips and more than 1.5 mm in 13 hips, 2 in the latter group apparently having worn through the polyethylene liner. The true wear may be twice as great. The use of 32 mm stainless steel heads and thin polyethylene inlays may have aggravated the wear problems. These HA-coated prostheses must all be checked regularly, so that cases with excessive polyethylene wear can be reoperated on before wear-through of the liner.Keywords
This publication has 30 references indexed in Scilit:
- Preliminary study of the effect of aging following irradiation on the wear of ultrahigh-molecular-weight polyethyleneThe Journal of Arthroplasty, 1995
- Severe polyethylene wear in total hip arthroplastyThe Journal of Arthroplasty, 1994
- Two-year outcome of hydroxyapatite-coated prostheses: Two femoral prostheses retrieved at autopsyActa Orthopaedica, 1994
- The shear strength and the failure mode of plasma‐sprayed hydroxyapatite coating to bone: The effect of coating thicknessJournal of Biomedical Materials Research, 1993
- Polyethylene wear in total hip arthroplasty in patient-matched groups: A comparison of stainless steel, cobalt chrome, and titanium-bearing surfacesThe Journal of Arthroplasty, 1993
- Osteolysis from a press-fit hydroxyapatite-coated implantThe Journal of Arthroplasty, 1993
- Hydroxyapatite-coated femoral stems. Histological analysis of components retrieved at autopsy.Journal of Bone and Joint Surgery, 1991
- Effect of femoral head size on wear of the polyethylene acetabular component.Journal of Bone and Joint Surgery, 1990