Catheter tip pH electrodes for continuous intravascular recording
- 1 January 1980
- journal article
- research article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 4 (3) , 122-124
- https://doi.org/10.3109/03091908009161104
Abstract
A small, flexible catheter-tip pH electrode has been developed for continuous measurement of intravascular pH in man and animals. The electrode utilises a pH sensitive polymer membrane rather than pH glass. The electrode passes through the lumen of a conventional cardiac catheter, by which it may be introduced to any desired site in the circulation. The electrode has no cross-sensitivity to other ions, to osmolality or oxygen tension. It is insensitive to flow and pressure. The response time is less than 300 ms, and the mean drift in vivo is 0.01 pH units/h. Continuous recordings of intravascular pH in animals have been made for up to Th. The electrodes can be sterilised, and have been used in man.Keywords
This publication has 4 references indexed in Scilit:
- CONTINUOUS MEASUREMENT OF pH IN CENTRAL ARTERIES AND VEINSThe Lancet, 1979
- A catheter-tip potassium-selective electrodeJournal of Medical Engineering & Technology, 1977
- Polymer membrane sensors for continuous intravascular monitoring of blood pHJournal of Applied Physiology, 1976
- Continuous blood pH monitoring using an indwelling catheterThe American Journal of Surgery, 1968