The effect of equipment set up on patient radiation dose in conventional and CT angiography of the renal arteries
- 1 September 2003
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 76 (909) , 625-630
- https://doi.org/10.1259/bjr/20434900
Abstract
Patient radiation dose in angiography of the renal arteries was assessed and optimized after installing new radiological equipment. In three separate studies (n=50, 25 and 20) patient exposure was monitored in detail. For the first study default factory settings were used, for the second the number of digital subtraction angiography (DSA) images was halved and the X-ray beam filtering during fluoroscopy was increased, and for the third study filtering during DSA was increased as well. Standard projections were derived and used in Monte Carlo simulations to derive dose conversion coefficients to calculate effective dose from the dose-area product (DAP). Dose conversion coefficients were also calculated for CT angiography (CTA). Using default factory settings on the new angiography system, DAP, number of images and effective dose were much higher than on the replaced unit. For the studies given above, DAP was reduced from 144 Gy cm(2) to 65 Gy cm(2) to 32 Gy cm(2), and effective dose from 22 mSv to 11 mSv to 9.1 mSv, respectively. Effective dose due to CTA was 5.2 mSv. It is concluded that modern angiography systems, resulting in high customer satisfaction, may readily cause much higher patient exposure than older systems. These doses may also be much higher than necessary. Optimization before putting such systems into use is absolutely essential. Internationally accepted recommendations for image quality and technique factors in angiography would be of great help.Keywords
This publication has 14 references indexed in Scilit:
- Patient and occupational dose in neurointerventional proceduresNeuroradiology, 2002
- Diagnostic Tests for Renal Artery Stenosis in Patients Suspected of Having Renovascular HypertensionAnnals of Internal Medicine, 2001
- Skin Injuries from Fluoroscopically Guided ProceduresAmerican Journal of Roentgenology, 2001
- Optimierung von Zusatzfilterung, Durchleuchtungskennlinie und Bildverstärker-Eingangsdosisleistung zur Reduktion der Strahlenexposition bei der AngiographieRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1999
- Diagnostic imaging of renal artery stenosisJournal of Human Hypertension, 1999
- Patient dosimetry in abdominal arteriographyPhysics in Medicine & Biology, 1999
- Patient dose values in interventional radiologyThe British Journal of Radiology, 1995
- Potentially higher patient radiation doses using digital equipment for barium studiesThe British Journal of Radiology, 1995
- Calculation of dose conversion factors for posterior–anterior chest radiography of adults with a relatively high-energy X-ray spectrumThe British Journal of Radiology, 1994
- Patient doses received during digital subtraction angiographyThe British Journal of Radiology, 1993