Left Ventricular Mechanical Assist Devices and Cardiac Device Interactions: An Observational Case Series
- 23 June 2009
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 32 (7) , 879-887
- https://doi.org/10.1111/j.1540-8159.2009.02403.x
Abstract
Nonpulsatile left ventricular assist devices (LVADs) are increasingly used for treatment of refractory heart failure. A majority of such patients have implanted cardiac devices, namely implantable cardioverter-defibrillators (ICDs) or cardiac resynchronization therapy-pacemaker (CRT-P) or cardiac resynchronization therapy-defibrillator (CRT-D) devices. However, potential interactions between LVADs and cardiac devices in this category of patients remain unknown. We reviewed case records and device logs of 15 patients with ICDs or CRT-P or CRT-D devices who subsequently had implantation of a VentrAssist LVAD (Ventracor Ltd., Chatswood, Australia) as destination therapy or bridge to heart transplantation. Pacemaker and ICD lead parameters before and after LVAD implant were compared. In addition, ventricular tachyarrhythmia event logs and potential electromagnetic interference reports were evaluated. Right ventricular (RV) sensing decreased in the first 6 months post-LVAD. Mean R-wave amplitude preimplant was 10.9 +/- 5.25 mV compared with 7.2 +/- 3.4 mV during follow-up (P = 0.02). RV impedance also decreased from 642 +/- 240 ohms at baseline to 580 +/- 212 ohms at follow-up (P = 0.007). There was a significant increase in RV stimulation threshold following implantation of the LVAD from 0.8 +/- 0.6 V at baseline to 1.4 +/- 1.0 V in the first 6 months postimplant (P = 0.01). A marked increase in ventricular tachyarrhythmia burden was observed in three patients. One patient displayed electromagnetic interference between the LVAD and defibrillator, resulting in inappropriate defibrillation therapy. LVADs have a definite impact on cardiac devices in respect with alteration of lead parameters, ventricular tachyarrhythmias, and electromagnetic interference.Keywords
This publication has 18 references indexed in Scilit:
- VentrAssist™ left ventricular assist device: clinical trial results and Clinical Development Plan updateEuropean Journal of Cardio-Thoracic Surgery, 2007
- Use of a Continuous-Flow Device in Patients Awaiting Heart TransplantationNew England Journal of Medicine, 2007
- First Clinical Implant of the VentrAssist Left Ventricular Assist System as Destination Therapy for End-Stage Heart FailureThe Journal of Heart and Lung Transplantation, 2005
- The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart FailureNew England Journal of Medicine, 2005
- Electrical and mechanical support in advanced heart failure. Rationale and feasibility of a combined management strategy.Published by Oxford University Press (OUP) ,2002
- Cardiac Resynchronization in Chronic Heart FailureNew England Journal of Medicine, 2002
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- Long-Term Use of a Left Ventricular Assist Device for End-Stage Heart FailureNew England Journal of Medicine, 2001
- A Randomized Study of the Prevention of Sudden Death in Patients with Coronary Artery DiseaseNew England Journal of Medicine, 1999
- Improved Survival with an Implanted Defibrillator in Patients with Coronary Disease at High Risk for Ventricular ArrhythmiaNew England Journal of Medicine, 1996