A phantom study on the positioning accuracy of the Novalis Body system
- 17 November 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (12) , 3052-3060
- https://doi.org/10.1118/1.1626122
Abstract
A phantom study was conducted to investigate inherent positioning accuracy of an image‐guided patient positioning system—the Novalis Body system for three‐dimensional (3‐D) conformal radiotherapy. This positioning system consists of two infrared (IR) cameras and one video camera and two kV x‐ray imaging devices. The initial patient setup was guided by the IR camera system and the target localization was accomplished using the kV x‐ray imaging system. In this study, the IR marker shift and phantom rotation were simulated and their effects on the positioning accuracy were examined by a Rando phantom. The effects of CT slice thickness and treatment sites on the positioning accuracy were tested. In addition, the internal target shift was simulated and its effect on the positioning accuracy was examined by a water tank. With the application of the Novalis Body system, the positioning error of the planned isocenter was significantly reduced. The experimental results with the simulated IR marker shifts indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.5±0.2, and 0.7±0.2 mm along the lateral, longitudinal, and vertical axes, respectively. The experimental results with the simulated phantom rotations indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.7±0.2, and 0.5±0.2 mm along the three axes, respectively. The experimental results with the simulated target shifts indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.7±0.2, and 0.5±0.2 mm along the three axes, respectively. On average, the positioning accuracy of 1 mm for the planned isocenter was achieved using the Novalis Body system.Keywords
This publication has 13 references indexed in Scilit:
- A technique of intensity‐modulated radiosurgery (IMRS) for spinal tumorsMedical Physics, 2002
- Clinical use of stereoscopic X-ray positioning of patients treated with conformal radiotherapy for prostate cancerInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Commissioning and quality assurance of an optically guided three-dimensional ultrasound target localization system for radiotherapyMedical Physics, 2002
- Daily prostate targeting using implanted radiopaque markersInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Fiducial‐based targeting accuracy for external‐beam radiotherapyMedical Physics, 2002
- Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibilityInternational Journal of Radiation Oncology*Biology*Physics, 2001
- Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancersInternational Journal of Radiation Oncology*Biology*Physics, 2000
- Commissioning an image-guided localization system for radiotherapyInternational Journal of Radiation Oncology*Biology*Physics, 2000
- Ultrasound-based stereotactic guidance of precision conformal external beam radiation therapy in clinically localized prostate cancerUrology, 2000
- Multi-modal volume registration by maximization of mutual informationMedical Image Analysis, 1996