Initial Clinical Experience with a New Small Sized Third-Generation Implantable Cardioverter Defibrillator: Results of a Multicenter Study
- 1 December 1997
- journal article
- clinical trial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 20 (12) , 2967-2974
- https://doi.org/10.1111/j.1540-8159.1997.tb05468.x
Abstract
This study reports the acute clinical experience with the new CPI VENTAK MINI: a small sized (68 cc), implantable cardioverter defibrillator (ICD) with 33 J stored energy. Implantation of the device was attempted in 113 patients (90 men, mean age 57 +/- 16 years, 64 with coronary artery disease, mean left ventricular ejection fraction 41%) with ventricular tachycardia or ventricular fibrillation (VF). All 113 patients (100%) were ultimately implanted, 12% of them for ICD replacement. Transvenous lead implantation was accomplished in all 104 patients (100%) receiving new leads, 95% of them with a single lead configuration. The safety criteria for implantation (2 consecutive VF conversions at 15 J or 3 at 20 J, in both cases without failures to convert) were demonstrated in all but 7 patients (6%). In 6 of these, safety criteria were not fully assessed while in the last patient defibrillation efficacy was not determined. Of the 104 patients with new leads, 90% underwent pectoral implantation. Of the 9 patients (9%) abdominally implanted, only 4 (4%) (3 children) were judged small sized for pectoral implant. At predischarge testing, reliable VF detection and conversion were noted in 96 of 97 patients tested. There was no perioperative mortality. At a 3.6 +/- 1.3 months follow-up, 34% of the patients had a spontaneous arrhythmic event, and 24% of the patients received shocks. Clinically inappropriate therapies occurred in 8% of the episodes in which any kind of therapy was delivered. This study demonstrates the short-term clinical efficacy and safety of the new device, and that pectoral implantation can be performed in the large majority of patients.Keywords
This publication has 22 references indexed in Scilit:
- Italian Multicenter Clinical Experience with Endocardial Defibrillation: Acute and Long‐Term Results in 307 PatientsPacing and Clinical Electrophysiology, 1995
- Optimizing Defibrillation Through Improved WaveformsPacing and Clinical Electrophysiology, 1995
- Cardioverter-defibrillator implantation in the catheterization laboratory: Initial experiences in 48 patientsAmerican Heart Journal, 1995
- A new transvenous internal cardioverter-defibrillator: Implantation technique, complications, and short-term follow-upAmerican Heart Journal, 1995
- Clinical Experience with a New Cardioverter Defibrillator Capable of Biphasic Waveform Pulse and Enhanced Data Storage: Results of a Prospective Multicenter StudyPacing and Clinical Electrophysiology, 1994
- Determinats of successful nonthoracotomy cardioverter-defibrillator implantation: Experience in 101 patients using two different lead systemsJournal of the American College of Cardiology, 1993
- Subpectoral implantation of cardioverter-defibrillator combined with a nonepicardial lead system: Preliminary experience with a novel approachThe American Journal of Cardiology, 1993
- Experience with a third-generation implantable cardioverter-defibrillatorThe American Journal of Cardiology, 1991
- Initial clinical experience with endocardial defibrillation using an implantable cardioverter/defibrillator with a triple-electrode systemArchives of internal medicine (1960), 1989
- Prospective comparison of sequential pulse and single pulse defibrillation with use of two different clinically available systemsJournal of the American College of Cardiology, 1989