A Continuous Glucose Sensor Based on Wired Enzyme™ Technology - Results from a 3-Day Trial in Patients with Type 1 Diabetes
- 1 October 2003
- journal article
- research article
- Published by Mary Ann Liebert Inc in Diabetes Technology & Therapeutics
- Vol. 5 (5) , 769-779
- https://doi.org/10.1089/152091503322526978
Abstract
Wired Enzyme™ (Therasense, Alameda, CA) sensing technology for continuous measurement of in vivo glucose concentrations offers the benefits of (1) excellent sensor stability, (2) reduced susceptibility to variations of in vivo oxygen concentration, and (3) minimized response to common electroactive interferents. This study describes the response of 48 sensors (25 in the upper arm, 23 in the abdomen) implanted for 3 days in patients with Type 1 diabetes. Prospective calibration was performed using capillary blood; results were compared with venous plasma glucose values obtained at 15-min intervals. Ninety-eight percent of readings fell in the clinically accurate Clarke error grid zone A or clinically acceptable zone B. Choice of the site of the implanted sensor (upper arm vs. abdomen) or the capillary blood calibration site (arm vs. finger) did not affect system accuracy.Keywords
This publication has 5 references indexed in Scilit:
- Alternate Site Glucose Testing: A Crossover DesignDiabetes Technology & Therapeutics, 2002
- High Incidence of Maternal Transmission of Diabetes in Obese Uruguayan ChildrenDiabetes Care, 2001
- Characterization of implantable biosensor membrane biofoulingAnalytical and Bioanalytical Chemistry, 2000
- Design and Optimization of a Selective Subcutaneously Implantable Glucose Electrode Based on "Wired" Glucose OxidaseAnalytical Chemistry, 1995
- The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes MellitusNew England Journal of Medicine, 1993