A method of obtaining the Awallcorrection factor of an ionization chamber in a60Co gamma -ray beam using tissue-air ratio data
- 1 March 1995
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 40 (3) , 383-391
- https://doi.org/10.1088/0031-9155/40/3/004
Abstract
A method is reported that evaluates the wall correction factor (Awall)water of an ionization chamber using measured data. The method needs a revised type of tissue-air ratio (TARcav) that should be measured with the ionization chamber. An experimental study has been carried out for 60Co gamma-rays using zero-area TARcav data extrapolated from non-zero-area TARcav data using the extrapolation method developed by Bjärngard and Petti. Using this technique, it is found that the value of (Awall)water for an ionization chamber has a deviation of -0.1% to -0.4% as compared with the Monte Carlo results. An experimental study has also been carried out using a published set of 60Co traditional tissue-air ratio (TAR) data assuming TARcav = TAR.Keywords
This publication has 10 references indexed in Scilit:
- Aeqand other factors in a60Co beam for a spherical or cylindrical minimal phantom and for an ionization chamber with a known Awallcorrection factorPhysics in Medicine & Biology, 1994
- A method of making contamination-free revised scatter-maximum ratios (SMRs) for 10 MV X-raysPhysics in Medicine & Biology, 1989
- Description of the scatter component in photon-beam dataPhysics in Medicine & Biology, 1988
- Monte Carlo calculations of scatter dose for small field sizes in a 6 0 Co beamMedical Physics, 1986
- A method of calculating high-energy photon primary absorbed dose in water using forward and backward spread dose-distribution functionsMedical Physics, 1985
- Ion chamber response and Awallcorrection factors in a60Co beam by Monte Carlo simulationPhysics in Medicine & Biology, 1985
- A simple derivation of N gas , a correction in A wall , and other comments on the AAPM Task Group 21 protocolMedical Physics, 1984
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Calculated response and wall correction factors for ionization chambers exposed to 60Co gamma‐raysMedical Physics, 1981
- Revision of tissue‐maximum ratio and scatter‐maximum ratio concepts for cobalt 60 and higher energy x‐ray beamsMedical Physics, 1980