Feasibility of automatic marker detection with an a-Si flat-panel imager
- 21 March 2001
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 46 (4) , 1219-1230
- https://doi.org/10.1088/0031-9155/46/4/321
Abstract
Here we study automatic detection of implanted gold markers relative to the field boundary in portal images for on-line position verification. Portal images containing 1-2 MU were taken with an amorphous silicon flat-panel imager. The images were obtained with lateral field at 18 MV. Both the detection success rate and the localization accuracy of markers of 1.0 and 1.2 mm diameter were determined with the help of a marker detection method based on a marker extraction kernel. A method for determining a fiducial reference point related to the field boundary was developed. Detection success rates were 0.99, 0.90 and 0.95 for markers of 1.2 mm diameter and 5 mm length, 1.0 mm diameter and 5 mm length and 1.0 mm diameter and 10 mm length respectively. The localization accuracy appeared to be better than 0.3 mm. The reference point could be reproduced with an accuracy equal to 1 pixel (0.5 mm at isocentre) within one fraction. During the first few seconds of a treatment fraction the field edge was not stable, which appeared to be an effect of the motion of the radiation source. Thanks to the use an a-Si flat-panel imager, on-line position verification using implanted gold markers becomes clinically feasible. We can use a clinically acceptable marker diameter as small as 1.0 mm. These markers can be automatically detected in portal images obtained with 1-2 MU relative to a stable reference point related to the field boundary.Keywords
This publication has 22 references indexed in Scilit:
- High-precision conformal radiotherapy (HPCRT) of prostate cancer—a new technique for exact positioning of the prostate at the time of treatmentInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imagingInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Prototype amorphous silicon array based radiotherapy portal imagerPublished by SPIE-Intl Soc Optical Eng ,1997
- Megavoltage imaging with a large-area, flat-panel, amorphous silicon imagerInternational Journal of Radiation Oncology*Biology*Physics, 1996
- Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibilityInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Prostate motion during standard radiotherapy as assessed by fiducial markersRadiotherapy and Oncology, 1995
- Measurement of prostate movement over the course of routine radiotherapy using implanted markersInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Methods for fully automated verification of patient set-up in external beam radiotherapy with polygon shaped fieldsPhysics in Medicine & Biology, 1994
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992
- Radiation field edge detection in portal imagesPhysics in Medicine & Biology, 1991