Volume delivery during high frequency oscillation
Open Access
- 1 March 1998
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood: Fetal & Neonatal
- Vol. 78 (2) , F148-F150
- https://doi.org/10.1136/fn.78.2.f148
Abstract
AIM To examine the delivered volume during “high volume strategy” high frequency oscillation, used as rescue treatment in preterm infants; and to identify factors, other than frequency and oscillatory amplitude, influencing the magnitude of volume delivery. METHOD Twenty infants (median gestational age 29 weeks) were studied on 45 occasions. Two oscillator types were used (SensorMedics and SLE). Delivered volume was measured under clinical conditions with the arterial blood gases within a predetermined range. A specially calibrated pneumotachograph system was used. RESULTS Overall, the median delivered volume was 2.4 ml/kg (range 1.0 to 3.6 ml/kg); on 32 occasions the delivered volume was greater than 2.0 ml/kg and on seven greater than 3.0 ml/kg. The delivered volume related significantly to disease severity; there was an inverse correlation between delivered volume and both the oxygenation index (OI) (r=−0.51) and AaDO2 (r=−0.54). CONCLUSION Delivered volume during HFO may, in certain infants, exceed the anatomical dead space, permitting some direct alveolar ventilation.Keywords
This publication has 17 references indexed in Scilit:
- Hydrogen passivation of iron-related hole traps in siliconJournal of Applied Physics, 1996
- Volume delivery during positive pressure inflation — relationship to spontaneous tidal volume of neonatesEarly Human Development, 1995
- The effect of frequency and mean airway pressure on volume delivery during high‐frequency oscillationPediatric Pulmonology, 1993
- Combined conventional ventilation with high frequency oscillation in neonatesEuropean Journal of Pediatrics, 1991
- Measurement of tidal volume using a pneumotachometer during high-frequency oscillationCritical Care Medicine, 1990
- Frequency, Tidal Volume, and Mean Airway Pressure Combinations that Provide Adequate Gas Exchange and Low Alveolar Pressure during High Frequency Oscillatory Ventilation in RabbitsPediatric Research, 1990
- Alveolar data in healthy, awake neonates during spontaneous ventilation: A preliminary investigationPediatric Pulmonology, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Effects of frequency, tidal volume, and lung volume on CO2 elimination in dogs by high frequency (2-30 Hz), low tidal volume ventilation.Journal of Clinical Investigation, 1981
- Ventilation by high-frequency oscillationJournal of Applied Physiology, 1980