Measurement of tidal volume using a pneumotachometer during high-frequency oscillation
- 1 June 1990
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 18 (6) , 651-653
- https://doi.org/10.1097/00003246-199006000-00015
Abstract
To assess the accuracy of a pneumotachometer (PN) for tidal volume (VT) measurements during high-frequency oscillation (HFO), we determined simultaneously VT using a PN and a full body plethysmograph (PL) in 12 rabbits. HFO was delivered with an oscillator at a frequency of 10 Hz, mean airway pressure of 8 cm H2O, and inspiratory time of 50%. Pressure amplitude (.DELTA. P) was varied as follows: 40, 60, 80, 20, 100, 40 cm H2O. Finally, in ten rabbits a spacer equal in dead space (VD) to that of the PN (15 ml) was left in-line for 5 min. Blood gases were obtained before and after the spacer was added. We found that VT-PN correlates well with VT-PL (r = .92), although the difference between VT-PN and VT-PL is greater at large. VT. Significant respiratory acidosis developed with the spacer in-line. PN may be used to trend VT during HFO but PN must not be left in-line, as increased VD seriously affects ventilation.Keywords
This publication has 7 references indexed in Scilit:
- Gas exchange in healthy rabbits during high-frequency oscillatory ventilationJournal of Applied Physiology, 1989
- Measurement of Tidal Volume during High Frequency Ventilation by Impedance PlethysmographyPediatric Research, 1988
- Effect of Distal Endotracheal Bias Flow on PaCO2 during High Frequency Oscillatory VentilationNeonatology, 1988
- Randomized trial of high-frequency jet ventilation versus conventional ventilation in respiratory distress syndromeThe Journal of Pediatrics, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Measurement of Tidal Volume during High- Frequency Jet Ventilation1Pediatric Research, 1985
- Effects of frequency, tidal volume, and lung volume on CO2 elimination in dogs by high frequency (2-30 Hz), low tidal volume ventilation.Journal of Clinical Investigation, 1981