A disintegration test for vaginal tablets: comparison with BP test
- 1 November 1990
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Pharmacy and Pharmacology
- Vol. 42 (11) , 795-796
- https://doi.org/10.1111/j.2042-7158.1990.tb07023.x
Abstract
To improve the disintegration test for vaginal tablets described in the British Pharmacopoeia (BP), a monitoring apparatus was added, and tested with seven effervescent and five non-effervescent tablets. A tablet was placed on the metal disc of the BP disintegration apparatus, and the guided plate was placed on the tablet. The guided plate moved downward smoothly in a cylinder as the tablet disintegrated. The movement was recorded by using a kymograph. The end-point and process of disintegration of the tablet were automatically recorded and the results obtained suggest that the modified test is a useful tool for the quality control of vaginal tablets.This publication has 1 reference indexed in Scilit:
- Design and Evaluation of a New Disintegration Test Apparatus for Vaginal TabletsYAKUGAKU ZASSHI, 1986