A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results
- 14 November 2001
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 46 (12) , 3191-3213
- https://doi.org/10.1088/0031-9155/46/12/309
Abstract
Methods based on magnetic resonance imaging for the measurement of three-dimensional distributions of radiation dose are highly developed. However, relatively little work has been done on optical computed tomography (OCT). This paper describes a new OCT scanner based on a broad beam light source and a two-dimensional charge-coupled device (CCD) detector. A number of key design features are discussed including the light source; the scanning tank, turntable and stepper motor control; the diffuser screen onto which images are projected and the detector. It is shown that the non-uniform pixel sensitivity of the low-cost CCD detector used and the granularity of the diffuser screen lead to a serious ring artefact in the reconstructed images. Methods are described for eliminating this. The problems arising from reflection and refraction at the walls of the gel container are explained. Optical ray-tracing simulations are presented for cylindrical containers with a variety of radii and verified experimentally. Small changes in the model parameters lead to large variations in the signal intensity observed in the projection data. The effect of imperfect containers on data quality is discussed and a method based on a 'correction scan' is shown to be successful in correcting many of the related image artefacts. The results of two tomography experiments are presented. In the first experiment, a radiochromic Fricke gel sample was exposed four times in different positions to a 100 kVp x-ray beam perpendicular to the plane of imaging. Images of absorbed dose with slice thickness of 140 μm were acquired, with 'true' in-plane resolution of 560 × 560 μm2 at the edge of the 72 mm field of view and correspondingly higher resolution at the centre. The nominal doses measured correlated well with the known exposure times. The second experiment demonstrated the well known phenomenon of diffusion in the dosemeter gels and yielded a value of (0.12 ± 0.02) mm2 s−1 for the diffusion coefficient of the xylenol orange/iron complex. Finally, the overall implications of the above findings for dosimetry using OCT are discussed.Keywords
This publication has 16 references indexed in Scilit:
- Radiation dosimetry using polymer gels: methods and applications.The British Journal of Radiology, 2000
- Polymer gel dosimetry using x-ray computed tomography: a feasibility study4Physics in Medicine & Biology, 2000
- Tissue-equivalent gel for non-invasive spatial radiation dose measurementsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2000
- Linear accelerator radiosurgery in the management of brain tumoursAnnals of Medicine, 2000
- Three-dimensional radiation dose measurements with Ferrous Benzoic Acid Xylenol Orange in Gelatin gel and optical absorption tomographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- Optical CT reconstruction of 3D dose distributions using the ferrous–benzoic–xylenol (FBX) gel dosimeterMedical Physics, 1998
- An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetryPhysics in Medicine & Biology, 1998
- Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gelPhysics in Medicine & Biology, 1996
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993
- Imaging of radiation dose by visible color development in ferrous‐agarose‐xylenol orange gelsMedical Physics, 1991