Penalized weighted least-squares image reconstruction for positron emission tomography
- 1 June 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 13 (2) , 290-300
- https://doi.org/10.1109/42.293921
Abstract
Presents an image reconstruction method for positron-emission tomography (PET) based on a penalized, weighted least-squares (PWLS) objective. For PET measurements that are precorrected for accidental coincidences, the author argues statistically that a least-squares objective function is as appropriate, if not more so, than the popular Poisson likelihood objective. The author proposes a simple data-based method for determining the weights that accounts for attenuation and detector efficiency. A nonnegative successive over-relaxation (+SOR) algorithm converges rapidly to the global minimum of the PWLS objective. Quantitative simulation results demonstrate that the bias/variance tradeoff of the PWLS+SOR method is comparable to the maximum-likelihood expectation-maximization (ML-EM) method (but with fewer iterations), and is improved relative to the conventional filtered backprojection (FBP) method. Qualitative results suggest that the streak artifacts common to the FBP method are nearly eliminated by the PWLS+SOR method, and indicate that the proposed method for weighting the measurements is a significant factor in the improvement over FBP.Keywords
This publication has 35 references indexed in Scilit:
- On complete-data spaces for PET reconstruction algorithmsIEEE Transactions on Nuclear Science, 1993
- Regularized emission image reconstruction using imperfect side informationIEEE Transactions on Nuclear Science, 1992
- Joint maximum likelihood estimation of emission and attenuation densities in PETPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1991
- On Bayesian image reconstruction from projections: uniform and nonuniform a priori source informationIEEE Transactions on Medical Imaging, 1989
- Enhanced Detection of Focal Brain Responses Using Intersubject Averaging and Change-Distribution Analysis of Subtracted PET ImagesJournal of Cerebral Blood Flow & Metabolism, 1988
- Physical performance of the latest generation of commercial positron scannerIEEE Transactions on Nuclear Science, 1988
- Noise and Edge Artifacts in Maximum-Likelihood Reconstructions for Emission TomographyIEEE Transactions on Medical Imaging, 1987
- The Use of Sieves to Stabilize Images Produced with the EM Algorithm for Emission TomographyIEEE Transactions on Nuclear Science, 1985
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982
- Nonparametric Maximum Likelihood Estimation by the Method of SievesThe Annals of Statistics, 1982