Influence of the needle bore size on platelet count and routine coagulation testing
- 1 October 2006
- journal article
- Published by Wolters Kluwer Health in Blood Coagulation & Fibrinolysis
- Vol. 17 (7) , 557-561
- https://doi.org/10.1097/01.mbc.0000245300.10387.ca
Abstract
The phlebotomy technique, particularly the use of small-bore needles, may influence the reliability of coagulation testing and platelet count. Routine coagulation tests were assayed in blood specimens collected from 22 consecutive patients in three separate, sequential phlebotomies, using butterfly devices with different needle sizes. Test results of samples collected with 23 and 25 G needles were compared with those obtained with the currently recommended 21 G needle. Although both the prothrombin time and activated partial thromboplastin time displayed a trend towards lower values employing the smaller 23 and 25 G needles, results did not differ significantly from the reference 21 G needle specimen, with the exceptions of D-dimer (25 G versus 21 G needle, 186 ± 70 versus 178 ± 66/ml, P < 0.01) and platelet count (23 G versus 21 G needle, 246 ± 55 versus 254 ± 56 × 10−3/l, P < 0.01; 25 G versus 21 G needle, 240 ± 55 versus 254 ± 56 × 10−3/l, P < 0.01). None of the mean biases recorded for the parameters was clinically meaningful, nor did they exceed the current desirable analytical quality specifications for desirable bias. Results of the present investigation suggest that, when a proper technique is used and within certain limitations, butterfly devices with small-bore needles may be a reliable alternative to draw venous blood for platelet count and coagulation testing.Keywords
This publication has 28 references indexed in Scilit:
- Interference of Blood Cell Lysis on Routine Coagulation TestingArchives of Pathology & Laboratory Medicine, 2006
- Preanalytical variability: the dark side of the moon in laboratory testingcclm, 2006
- Short-term venous stasis influences routine coagulation testingBlood Coagulation & Fibrinolysis, 2005
- No influence of a butterfly device on routine coagulation assays and D‐dimer measurementJournal of Thrombosis and Haemostasis, 2005
- A shortened activated partial thromboplastin time is associated with the risk of venous thromboembolismBlood, 2004
- Effect of Specimen Collection on Routine Coagulation Assays and D-Dimer MeasurementClinical Chemistry, 2004
- Screening for the Risk for Bleeding or ThrombosisAnnals of Internal Medicine, 2003
- Phlebotomy Tools of the TradeHome Healthcare Now, 2002
- Preanalytical aspects of routine coagulation measurementsScandinavian Journal of Clinical and Laboratory Investigation, 2001
- Does transfusion using a syringe infusion pump and small‐gauge needle cause hemolysis?Transfusion, 1981