Noninvasive cardiac output monitoring (NICOM): a clinical validation
Top Cited Papers
- 26 April 2007
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 33 (7) , 1191-1194
- https://doi.org/10.1007/s00134-007-0640-0
Abstract
To evaluate the clinical utility of a new device for continuous noninvasive cardiac output monitoring (NICOM) based on chest bio-reactance compared with cardiac output measured semi-continuously by thermodilution using a pulmonary artery catheter (PAC-CCO). Prospective, single-center study. Intensive care unit. Consecutive adult patients immediately after cardiac surgery. Cardiac output measurements obtained from NICOM and thermodilution were simultaneously recorded minute by minute and compared in 110 patients. We evaluated the accuracy, precision, responsiveness, and reliability of NICOM for detecting cardiac output changes. Tolerance for each of these parameters was specified prospectively. A total of 65,888 pairs of cardiac output measurements were collected. Mean reference values for cardiac output ranged from 2.79 to 9.27 l/min. During periods of stable PAC-CCO (slope < ± 10%, 2SD/mean < 20%), the correlation between NICOM and thermodilution was R = 0.82; bias was +0.16 ± 0.52 l/min (+4.0 ± 11.3%), and relative error was 9.1% ± 7.8%. In 85% of patients the relative error was < 20%. During periods of increasing output, slopes were similar with the two methods in 96% of patients and intra-class correlation was positive in 96%. Corresponding values during periods of decreasing output were 90% and 84%, respectively. Precision was always better with NICOM than with thermodilution. During hemodynamic challenges, changes were 3.1 ± 3.8 min faster with NICOM (p < 0.01) and amplitude of changes did not differ significantly. Finally, sensitivity of the NICOM for detecting significant directional changes was 93% and specificity was 93%. Cardiac output measured by NICOM had most often acceptable accuracy, precision, and responsiveness in a wide range of circulatory situations.Keywords
This publication has 21 references indexed in Scilit:
- Non-invasive measurement of cardiac output in patients with chronic heart failureBlood Pressure Monitoring, 2004
- Estimation of left ventricular systolic function by single transpulmonary thermodilutionIntensive Care Medicine, 2004
- Reproducibility of thermodilution cardiac output determination in critically ill patients: Comparison between bolus and continuous methodJournal of Clinical Monitoring and Computing, 1996
- Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patientsCritical Care Medicine, 1995
- Cardiac output measurement by the thermodilution method: An in vitro test of accuracy of three commercially available automatic cardiac output computersIntensive Care Medicine, 1995
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Continuous noninvasive cardiac output as estimated from the pulse contour curveJournal of Clinical Monitoring and Computing, 1992
- Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedanceCritical Care Medicine, 1986
- Analysis of factors affecting the variability of fick versus indicator dilution measurements of cardiac outputThe American Journal of Cardiology, 1985