Inter-observer reproducibility of measurements of range of motion in patients with shoulder pain using a digital inclinometer
Open Access
- 14 June 2004
- journal article
- research article
- Published by Springer Nature in BMC Musculoskeletal Disorders
- Vol. 5 (1) , 18
- https://doi.org/10.1186/1471-2474-5-18
Abstract
Reproducible measurements of the range of motion are an important prerequisite for the interpretation of study results. The digital inclinometer is considered to be a useful instrument because it is inexpensive and easy to use. No previous study assessed inter-observer reproducibility of range of motion measurements with a digital inclinometer by physical therapists in a large sample of patients. Two physical therapists independently measured the passive range of motion of the glenohumeral abduction and the external rotation in 155 patients with shoulder pain. Agreement was quantified by calculation of the mean differences between the observers and the standard deviation (SD) of this difference and the limits of agreement, defined as the mean difference ± 1.96*SD of this difference. Reliability was quantified by means of the intraclass correlation coefficient (ICC). The limits of agreement were 0.8 ± 19.6 for glenohumeral abduction and -4.6 ± 18.8 for external rotation (affected side) and quite similar for the contralateral side and the differences between sides. The percentage agreement within 10° for these measurements were 72% and 70% respectively. The ICC ranged from 0.28 to 0.90 (0.83 and 0.90 for the affected side). The inter-observer agreement was found to be poor. If individual patients are assessed by two different observers, differences in range of motion of less than 20–25 degrees can not be distuinguished from measurement error. In contrast, acceptable reliability was found for the inclinometric measurements of the affected side and the differences between the sides, indicating that the inclimeter can be used in studies in which groups are compared.Keywords
This publication has 29 references indexed in Scilit:
- Observer Reliability and AgreementPublished by Wiley ,2005
- How reliably do rheumatologists measure shoulder movement?Annals of the Rheumatic Diseases, 2002
- Intratester and intertester reliability of the Cybex electronic digital inclinometer (EDI-320) for measurement of active neck flexion and extension in healthy subjectsManual Therapy, 2001
- Variation in measurements of range of motion: a study in reflex sympathetic dystrophy patientsClinical Rehabilitation, 1998
- The efficacy of non-steroidal anti-inflammatory drugs (NSAIDS) for shoulder complaints. A systematic reviewJournal of Clinical Epidemiology, 1995
- Musculoskeletal disorders and disability in persons aged 85 and over: a community survey.Annals of the Rheumatic Diseases, 1994
- OBSERVER VARIABILITY IN MEASURING ELEVATION AND EXTERNAL ROTATION OF THE SHOULDERRheumatology, 1994
- Ranges of movement at the shoulder complex using an electromagnetic movement sensor.Annals of the Rheumatic Diseases, 1991
- Clinical Reliability of Shoulder Function Assessment in Patients with Rheumatoid ArthritisScandinavian Journal of Rheumatology, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986