Direct Determination of Volatile Elements in Nickel Alloys by Electrothermal Vaporization Inductively Coupled Plasma Mass Spectrometry
- 1 January 1997
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 12 (2) , 131-135
- https://doi.org/10.1039/a603313j
Abstract
A method is described for the direct determination of Bi, Pb and Te in solid Ni alloys by ETV-ICP-MS. Samples are introduced into the graphite tube as small filings or chips weighing up to 3 mg. Using diluted sea water as a physical carrier, both Bi and Pb could be determined in solid Ni using external calibration with aqueous samples although results for Pb were biased low. Better results in terms of accuracy and precision were obtained when solid RMs (Ni) were used for calibration. LODs of 14 and 44 ng g -1 were obtained for Bi and Pb, respectively, using a reduced sensitivity mode (OmniRange). Based on signals obtained for solution standards measured at the highest sensitivity, LODs of 0.002 and 0.004 ng g -1 are possible for Bi and Pb, respectively. The determination of Te by this technique was not successful using either solution or solids calibration. Tellurium did not show a linear instrument response with concentration, which was probably due to an interaction between the Te and one or more matrix components in the solid phase that alters the release mechanism(s) for Te from those observed for Pb and Bi.Keywords
This publication has 17 references indexed in Scilit:
- Analyte Transport Studies of Aqueous Solutions and Slurry Samples Using Electrothermal Vaporization ICP-MSApplied Spectroscopy, 1995
- Use of the Ar2+signal as a diagnostic tool in solid sampling electrothermal vaporization inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1995
- Solid sampling electrothermal vaporization inductively coupled plasma mass spectrometry for the determination of arsenic in standard reference materials of plant originJournal of Analytical Atomic Spectrometry, 1995
- Direct analysis of solids by ultrasonic slurry electrothermal vaporization inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1994
- Determination of tellurium and antimony in nickel alloys by laser excited atomic fluorescence spectrometry in a graphite furnaceSpectrochimica Acta Part B: Atomic Spectroscopy, 1993
- Analysis of solid samples by ICP-mass spectrometryAnalytical and Bioanalytical Chemistry, 1992
- Direct solid sampling of nickel based alloys by graphite furnace atomic absorption spectrometry with aqueous calibrationSpectrochimica Acta Part B: Atomic Spectroscopy, 1990
- Determination of trace elements in metals by atomic-absorption spectrometry with the introduction of solid samples into furnaces: an investigation involving aqueous standardsMicrochimica Acta, 1982
- Determination of lead, bismuth, zinc, silver and antimony in steel and nickel-base alloys by atomic-absorption spectrometry using direct atomisation of solid samples in a graphite furnaceThe Analyst, 1979
- Atomic Absorption Determination of Lead, Bismuth, Selenium, Tellurium, Thallium and Tin in Complex Alloys Using Direct Atomization from Metal Chips in the Graphite FurnaceApplied Spectroscopy, 1977