Maximum-likelihood expectation-maximization reconstruction of sinograms with arbitrary noise distribution using NEC-transformations
- 1 May 2001
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 20 (5) , 365-375
- https://doi.org/10.1109/42.925290
Abstract
The maximum-likelihood (ML) expectation-maximization (EM) [ML-EM] algorithm is being widely used for image reconstruction in positron emission tomography. The algorithm is strictly valid if the data are Poisson distributed. However, it is also often applied to processed sinograms that do not meet this requirement. This may sometimes lead to suboptimal results: streak artifacts appear and the algorithm converges toward a lower likelihood value. As a remedy, the authors propose two simple pixel-by-pixel methods [noise equivalent counts (NEC)-scaling and NEC-shifting] in order to transform arbitrary sinogram noise into noise which is approximately Poisson distributed (the first and second moments of the distribution match those of the Poisson distribution). The convergence speed associated with both transformation methods is compared, and the NEC-scaling method is validated with both simulations and clinical data. These new methods extend the ML-EM algorithm to a general purpose nonnegative reconstruction algorithm.Keywords
This publication has 16 references indexed in Scilit:
- Attenuation correction in whole-body FDG oncological studies: the role of statistical reconstruction.European Journal of Nuclear Medicine and Molecular Imaging, 1999
- Statistical image reconstruction methods for randoms-precorrected PET scansMedical Image Analysis, 1998
- Iterative Reconstruction of Transmission Sinograms with Low Signal to Noise RatioPublished by Springer Nature ,1997
- Optimizing rod window width in positron emission tomographyIEEE Transactions on Medical Imaging, 1995
- Penalized weighted least-squares image reconstruction for positron emission tomographyIEEE Transactions on Medical Imaging, 1994
- The ECAT EXACT HRJournal of Computer Assisted Tomography, 1994
- Corrections for accidental coincidences and attenuation in maximum-likelihood image reconstruction for positron-emission tomographyIEEE Transactions on Medical Imaging, 1991
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Stopping Rule for the MLE Algorithm Based on Statistical Hypothesis TestingIEEE Transactions on Medical Imaging, 1987
- Accelerated Iterative Reconstruction for Positron Emission Tomography Based on the EM Algorithm for Maximum Likelihood EstimationIEEE Transactions on Medical Imaging, 1986