Abstract
Polyethylene has been used for more than 30 years as an orthopaedic bearing material; however, recently concern has been focused on the early failure of some polyethylene bearings. The damage seen in some bearings has been linked to gamma radiation sterilization performed in an air environment. Gamma sterilization in air has been documented to cause an increase in oxidation and degradation of mechanical properties that continue with time. However, not all retrieved bearings that are gamma sterilized in air exhibit the elevated oxidation and mechanical property degradation that lead to early component failure. Bearings that are gamma sterilized in air oxidize while sitting in inventory before implantation. Shelf oxidation rate was estimated based on analysis of a series of never implanted tibial bearings. This shelf oxidation rate allowed estimation of in vivo oxidation for retrieved tibial bearings of known sterilization date. Bearings with less than 1 year of shelf life after gamma sterilization in air had lower in vivo oxidation and better in vivo performance than did those with longer shelf life before implantation. Shelf time before implantation appears to be a significant factor in the success or failure of bearings that are gamma sterilized in air.Keywords
This publication has 11 references indexed in Scilit:
- Impact of gamma sterilization on clinical performance of polyethylene in the kneeThe Journal of Arthroplasty, 1996
- Challenge to the concept that UHMWPE acetabular components oxidize in vivoBiomaterials, 1993
- Mechanism of ageing in irradiated polymersPolymer Degradation and Stability, 1988
- The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement.Journal of Bone and Joint Surgery, 1986
- The effect of sterilising radiation on the properties of ultra-high molecular weight polyethyleneMaterials Chemistry and Physics, 1986
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983
- Aspects of Free Radical Reactions in Biological Systems: AgingJournal of Gerontology, 1980
- The effects of radiation sterilization on the properties of ultrahigh molecular weight polyethyleneJournal of Biomedical Materials Research, 1979
- Free Radicals and Inflammation: Protection of Synovial Fluid by Superoxide DismutaseScience, 1974
- Autoxidation of semicrystalline polyethylenePolymer Engineering & Science, 1966