Size Analysis of a Pressurized Metered Dose Inhaler–Delivered Suspension Formulation by the API Aerosizer Time-of-Flight Aerodynamic Particle Size Analyzer
- 1 January 1999
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine
- Vol. 12 (4) , 255-264
- https://doi.org/10.1089/jam.1999.12.255
Abstract
Aerosizer time-of-flight (TOF) aerodynamic particle size analyzers (TSI-Amherst, Amherst, MA) are widely used for the rapid assessment of aerosols from a wide variety of drug delivery devices, including pressurized metered dose inhalers (pMDIs). This technique offers significant advantages in terms of rapid measurement times in comparison with the more time-consuming compendial methods such as the cascade impactor or multistage liquid impinger. Particle size analysis takes place by determining the TOF of individual particles following acceleration to supersonic velocity. No drug assay is performed; thus, the resulting size distribution also includes particles that do not contain any medication such as the excipients and surfactant that are present in most pMDI-based formulations. Illustrative data are presented for one particular formulation (Pulmicort: 200 μg of budesonide per dose; Astra Draco; Lund, Sweden) and demonstrate that bias from this source can significantly shift the reported particle distribution to finer sizes compared with impactor-based analysis in which direct assay for drug has taken place. In this case, the mass median aerodynamic diameter (MMAD) determined by an Aerosizer-LD was close to 2.4 μm, but was found to be approximately 4 μm using the cascade impactor–based procedure. Such a shift results in an overestimation of the fine particle fraction of the emitted dose, which may lead to misleading conclusions about the therapeutic benefit of a particular drug delivery system when making use of this formulation. TOF aerosol measurement techniques appear to be vulnerable to this type of bias for any suspension formulation in which the drug content is not homogeneously distributed within all particle sizes.Keywords
This publication has 7 references indexed in Scilit:
- USE OF THE AEROSIZER® AERODYNAMIC PARTICLE SIZE ANALYZER TO CHARACTERIZE AEROSOLS FROM PRESSURIZED METERED-DOSE INHALERS (pMDIs) FOR MEDICATION DELIVERYJournal of Aerosol Science, 1999
- Time-of-Flight Aerodynamic Particle Size Analyzers: Their Use and Limitations for the Evaluation of Medical AerosolsJournal of Aerosol Medicine, 1999
- Inhalational drug delivery from seven different spacer devices.Thorax, 1996
- Choosing the metered-dose inhaler spacer or holding chamber that matches the patient's need: Evidence that the specific drug being delivered is an important considerationJournal of Allergy and Clinical Immunology, 1995
- Calibration and performance of an API AerosizerJournal of Aerosol Science, 1993
- Fraction of dose exhaled after administration of pressurized inhalation aerosolsInternational Journal of Pharmaceutics, 1980
- Aerosol Beam SpectrometryNature Physical Science, 1973