Fast T1 mapping on a whole-body scanner
Open Access
- 8 July 1999
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 42 (1) , 206-209
- https://doi.org/10.1002/(sici)1522-2594(199907)42:1<206::aid-mrm28>3.0.co;2-q
Abstract
A method for the fast acquisition of quantitative T1 maps is presented. It is based on the acquisition of a series of snapshot fast low‐angle shot (FLASH) images after inversion of the magnetization. A drawback of this method is the necessity for a sufficiently long relaxation delay before each inversion pulse, leading to long experimental times if averaging or segmentation is required. Thus implementation of this method on a whole‐body scanner is problematic, as the longer gradient rise times provide prolonged repetition times, which makes segmentation indispensable to provide a sufficient temporal resolution. We present a modification of the method that allows for reduction in the intermediate relaxation delays and thus for considerably reduced experimental durations. Magn Reson Med 42:206–209, 1999.Keywords