Liquid Helium Cryostat for Mössbauer Scattering Experiments
- 1 November 1965
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 36 (11) , 1607-1610
- https://doi.org/10.1063/1.1719402
Abstract
A simply constructed cryostat designed for Mössbauer γ-ray scattering experiments is described. The sample volume is part of the transfer line from a liquid helium reservoir to a mechanical pump; the sample is thus cooled by helium flow. The helium vapor exiting from the sample volume cools the transfer line and radiation shield. The temperature of the sample is controlled by the helium flow rate determined by a needle valve at the pump. The operating range is from 9 to ∼300°K. The liquid helium consumption in liters/h is 0.3 at 10°K, 0.2 at 16°K, and 0.1 at 34°K. Possible modifications to improve the operating characteristics are discussed.Keywords
This publication has 4 references indexed in Scilit:
- Mössbauer Scattering: The Gyromagnetic Moments of First States in and and andPhysical Review Letters, 1965
- Kryotechnik ‐ Experimentieren bei tiefen TemperaturenChemie Ingenieur Technik - CIT, 1964
- High Velocity Drive for Mössbauer ExperimentsReview of Scientific Instruments, 1963
- An Efficient Cryostat for Producing Temperatures Between 4° and 80°K: The Production of Liquid Hydrogen Targets Using Liquid HeliumReview of Scientific Instruments, 1954