Comparison of Plasma Total Homocysteine Measurements in 14 Laboratories: An International Study
Open Access
- 1 August 1999
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 45 (8) , 1261-1268
- https://doi.org/10.1093/clinchem/45.8.1261
Abstract
Background: Information on interlaboratory variation and especially on methodological differences for plasma total homocysteine is lacking. Methods: We studied 14 laboratories that used eight different method types: HPLC with electrochemical detection (HPLC-ED); HPLC with fluorescence detection (HPLC-FD) further subdivided by type of reducing/derivatizing agent; gas chromatography/mass spectrometry (GC/MS); enzyme immunoassay (EIA); and fluorescence polarization immunoassay (FPIA). Three of these laboratories used two methods. The laboratories participated in a 2-day analysis of 46 plasma samples, 4 additional plasma samples with added homocystine, and 3 plasma quality-control (QC) pools. Results were analyzed for imprecision, recovery, and methodological differences. Results: The mean among-laboratory and among-run within-laboratory imprecision (CV) was 9.3% and 5.6% for plasma samples, 8.8% and 4.9% for samples with added homocystine, and 7.6% and 4.2% for the QC pools, respectively. Difference plots showed values systematically higher than GC/MS for HPLC-ED, HPLC-FD using sodium borohydride/monobromobimane (however, for only one laboratory), and EIA, and lower values for HPLC-FD using trialkylphosphine/4-(aminosulfonyl)-7-fluoro-2,1,3-benzoxadiazole. The two HPLC-FD methods using tris(2-carboxyethyl) phosphine/ammonium 7-fluoro-2,1,3-benzoxadiazole-4-sulfonate (SBD-F) or tributyl phosphine/SBD-F, and the FPIA method showed no detectable systematic difference from GC/MS. Conclusions: Among-laboratory variations within one method can exceed among-method variations. Some of the methods tested could be used interchangeably, but there is an urgent need to improve analytical imprecision and to decrease differences among methods.Keywords
This publication has 19 references indexed in Scilit:
- Analytical Performance Characteristics Should Be Judged against Objective Quality SpecificationsClinical Chemistry, 1999
- Rapid and Accurate HPLC Assay for Plasma Total Homocysteine and Cysteine in a Clinical Laboratory SettingClinical Chemistry, 1999
- Method Comparison for Total Plasma Homocysteine between the Abbott IMx Analyzer and an HPLC Assay with Internal StandardizationClinical Chemistry, 1999
- Enzyme conversion immunoassay for determining total homocysteine in plasma or serumClinical Chemistry, 1998
- A Quantitative Assessment of Plasma Homocysteine as a Risk Factor for Vascular DiseaseJAMA, 1995
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- Homocyst(e)ine and arterial occlusive diseasesJournal of Internal Medicine, 1994
- Hyperhomocyst(e)inemia as a Risk Factor for Occlusive Vascular DiseaseAnnual Review of Nutrition, 1992
- Rapid high-performance liquid chromatographic assay for total homocysteine levels in human serumJournal of Chromatography B: Biomedical Sciences and Applications, 1991
- Homocyst(e)inemia in daily practiceCoronary Artery Disease, 1990