What design factors influence wear behavior at the bearing surfaces in total joint replacements?
- 1 January 2008
- journal article
- Published by Wolters Kluwer Health in Journal of the American Academy of Orthopaedic Surgeons
- Vol. 16, S101-S106
- https://doi.org/10.5435/00124635-200800001-00020
Abstract
Bearing surface wear in total joint replacements arises from local stresses that exceed the mechanical strength of the articulating materials. Because both the tensile/compressive principal stresses and maximum shear stress near the bearing surface increase when contact stresses increase, minimizing contact stresses has been a central design goal, especially in total knees. Wear rates increase with factors such as increased sliding distance in metal-on-polyethylene bearings, or suboptimal fluid film lubrication in the case of hard-on-hard total hip implants. These factors in turn depend directly on implant design. Advanced preclinical assessment technologies such as laboratory physical simulators and finite element analyses have provided means by which the dependence of wear rate on mechanical design factors can be quantified. However, untoward complexities occurring in vivo, such as impingement or third-body challenge, can appreciably compromise wear performance even for implants that are well-designed in terms of bearing surface stress minimization.Keywords
This publication has 17 references indexed in Scilit:
- Effects of implant design parameters on fluid convection, potentiating third-body debris ingress into the bearing surface during THA impingement/subluxationJournal of Biomechanics, 2007
- Association of Third Body Embedment with Rim Damage in Retrieved Acetabular LinersClinical Orthopaedics and Related Research, 2007
- Surface damage to an Oxinium femoral head prosthesis after dislocationThe Journal of Bone and Joint Surgery. British volume, 2007
- PRESIDENTIAL GUEST LECTURE: Tribology of Alternative BearingsClinical Orthopaedics and Related Research, 2006
- Metal-on-metal hip joint tribologyProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2006
- Acetabular Cup Geometry and Bone-Implant Interference have More Influence on Initial Periprosthetic Joint Space than Joint Loading and Surgical Cup InsertionJournal of Biomechanical Engineering, 2005
- Molecular chain stretch is a multiaxial failure criterion for conventional and highly crosslinked UHMWPEJournal of Orthopaedic Research, 2005
- Finite Element Analysis of Acetabular WearPublished by Wolters Kluwer Health ,1997
- Adaptiive finite element modeling of long‐term polyethylene wear in total hip arthroplastyJournal of Orthopaedic Research, 1996
- A sliding-distance-coupled finite element formulation for polyethylene wear in total hip arthroplastyJournal of Biomechanics, 1996