Human monocyte/macrophage response to cobalt‐chromium corrosion products and titanium particles in patients with total joint replacements
- 1 January 1997
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 15 (1) , 40-49
- https://doi.org/10.1002/jor.1100150107
Abstract
The responses of human peripheral blood monocytes of 10 normal volounteers and 14 patients with total hip replacements to particles of commercially pure titanium and chromium orthphosphate (a corrosion product from cobalt-chromium alloy implants) were studied. In addition, these phagocytosable particles were added to cultured mononuclear cells isolated from the interfacial membrane of 14 patients with failed implants. Peripheral blood monocytes from patients who had had a total hip replacement produced significantly higher levels of interleukin-1 (both interleukin-1α and interleukin-1β) and prostaglandin E2 follwing particulate stimulation than those from normal volunteers. Supernatants from both titanium and chromium orthophosphate-stimulated peripheral blood monocytes from the volunteers and patients with total hip replacement induced bone resorption (assayed in organ cultures of newborn mouse calvariae)and the proliferation of human fibroblasts. The levels of bone resorption were significantly higher (p < 0.05) in patients with implants that in normal volunteers. There were no significant differences in the responses of cells between patients with focal osteolysis and those without osteolysis. Interfacial membrane mononuclear cells also produced high levels of interleukin-1α, interleukin-1β, and prostaglandin E2 and expressed bone resorptive activities following stimulation with either titanium or chromium orthophosphate. More importantly, interfacial membrane mononuclear cells “spontaneously” produced high levels of prostaglandin E2 that were comparable with the response of peripheral blood monocytes stimulated with particulate wear debris. The clinical relevance of this study is 2-fold. First, mononuclear cells from patients with total hip replacement were some-how “sensitized” to metal particles in comparison with mononuclear cells from individuals without an implant. Second, the chromium orthophosphate corrosion product was a potent macrophage/monocyte activator and may contribute to macrophage-mediated osteolysis and aseptic loosening.Keywords
This publication has 35 references indexed in Scilit:
- PARTICULATE-INDUCED, PROSTAGLANDIN- AND CYTOKINE-MEDIATED BONE RESORPTION IN AN EXPERIMENTAL SYSTEM AND IN FAILED JOINT REPLACEMENTSClinical Journal of Sport Medicine, 1996
- Human monocyte response to particulate biomaterials generated in vivo and in vitroJournal of Orthopaedic Research, 1995
- The potential role of fibroblasts in periprosthetic osteolysis: Fibroblast response to titanium particlesJournal of Bone and Mineral Research, 1995
- Discoordinate expression of IL-1α and IL-1β in interfacial membranes of failed joint prosthesesInflammation Research, 1994
- In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy couplingJournal of Biomedical Materials Research, 1993
- Bone resorption activity of particulate-stimulated macrophagesJournal of Bone and Mineral Research, 1993
- Particulate-Associated Endosteal Osteolysis in Titanium-Base Alloy Cementless Total Hip ReplacementPublished by ASTM International ,1992
- Studies on IL-1 receptors on D10S T-helper cells: Demonstration of two molecularly and antigenically distinct IL-1 binding proteinsCytokine, 1989
- A Clinical-Pathologic-Biochemical Study of the Membrane Surrounding Loosened and Nonloosened Total Hip ArthroplastiesClinical Orthopaedics and Related Research, 1989
- Formation of a synovial‐like membrane at the bone‐cement interface: Its role in bone resorption and implant loosening after total hip replacementArthritis & Rheumatism, 1986