Calibration at Different Flow Rates of a Multistage Liquid Impinger
Open Access
- 1 January 1997
- journal article
- research article
- Published by Taylor & Francis in Aerosol Science and Technology
- Vol. 27 (1) , 39-49
- https://doi.org/10.1080/02786829708965456
Abstract
The multistage liquid impinger (MLI) is an impactor suitable for size-selective characterization and testing of pharmaceutical inhalation products, especially dry powder inhalers (DPIs). The four-stage version of the MLI, which was adopted by the European Pharmacopoeia for testing at 60 L/min, has been calibrated by two different laboratories (Astra Draco AB, Sweden, and AEA Technology, UK), reaching consistent results. The MLI has been extended with an additional impaction stage and all impaction stages have been calibrated at flow rates of 30, 40, 50, 60, 80, and 100 L/min. It is shown that in this flow range, the cutoff diameter depends on the flow rate according to a simplified impaction equation, i.e., the cutoff diameter is proportional to the inverse of the square root of the flow rate. Based on the calibration data and the simplified equation, a cutoff value for any flow rate in the range 30–100 L/min can be assigned. This is an important possibility as a DPI should be tested at a flow rate derived from the air flow resistance of the DPI. The cutoff values at 60 L/min for the four impaction stages are 13, 6.8, 3.1, and 1.7 μm, the fifth stage being a filter stage of the five-stage MLI. The apparatus has also been shown to be robust and to have a high degree of uniformity between replicas.Keywords
This publication has 9 references indexed in Scilit:
- The calibration of the astra draco multi-stage liquid impinger (MLI)Journal of Aerosol Science, 1995
- Experimental Verification of Using Cumulative Efficiency Curves for Determining Stage Constants of a Cascade ImpacterAerosol Science and Technology, 1995
- A Model for the Effect of Inhalation Device Flow Resistance on the Peak Inspiratory Flow Rate and Its Application in Pharmaceutical TestingJournal of Aerosol Medicine, 1994
- The Relationship Between Powder Inhaler Resistance and Peak Inspiratory Conditions in Healthy Volunteers — Implications for In Vitro TestingJournal of Aerosol Medicine, 1993
- A technique for assessing size-selective dust samplers using the APS and polydisperse test aerosolsJournal of Aerosol Science, 1991
- An assessment of an andersen mark-II cascade impactorJournal of Aerosol Science, 1988
- Calibration and Use of the Aerodynamic Particle Sizer (APS 3300)Aerosol Science and Technology, 1986
- Lorazepam on visuo-motor co-ordination and visual function in manJournal of Pharmacy and Pharmacology, 1973
- Multistage liquid impinger.Microbiology and Molecular Biology Reviews, 1966