Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy – a manikin study*
- 28 March 2006
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 61 (5) , 469-477
- https://doi.org/10.1111/j.1365-2044.2006.04547.x
Abstract
The Airtraq Laryngoscope is a novel intubation device which allows visualisation of the vocal cords without alignment of the oral, pharyngeal and tracheal axes. We compared the Airtraq with the Macintosh laryngoscope in simulated easy and difficult laryngoscopy. Twenty-five anaesthetists were allowed up to three attempts to intubate the trachea in each of three laryngoscopy scenarios using a Laerdal Intubation Trainer followed by five scenarios using a Laerdal SimMan Manikin. Each anaesthetist then performed tracheal intubation of the normal airway a second time to characterise the learning curve. In the simulated easy laryngoscopy scenarios, there was no difference between the Airtraq and the Macintosh in success of tracheal intubation. The time taken to intubate at the end of the protocol was significantly lower using the Airtraq (9.5 (6.7) vs. 14.2 (7.4) s), demonstrating a rapid acquisition of skills. In the simulated difficult laryngoscopy scenarios, the Airtraq was more successful in achieving tracheal intubation, required less time to intubate successfully, caused less dental trauma, and was considered by the anaesthetists to be easier to use.Keywords
This publication has 30 references indexed in Scilit:
- Lymphocyte contamination of laryngoscope blades – a possible vector for transmission of variant Creutzfeldt‐Jakob diseaseAnaesthesia, 2005
- A comparison of 20 laryngoscope blades using an intubating manikin: visual analogue scores and forces exerted during laryngoscopy*Anaesthesia, 2005
- Das EasyScope - Eine Alternative zum bewährten Macintosh-Laryngoskop?AINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie, 2004
- Manikin training for neonatal resuscitation with the laryngeal mask airwayPediatric Anesthesia, 2004
- Laryngoscopic IntubationAnesthesiology, 2003
- Presence of protein deposits on ‘cleaned’ re‐usable anaesthetic equipmentAnaesthesia, 2001
- The American Society of Anesthesiologists Closed Claims ProjectAnesthesiology, 1999
- A new variant of Creutzfeldt-Jakob disease in the UKPublished by Elsevier ,1996
- Predicting Difficult Endotracheal Intubation in Surgical Patients Scheduled for General AnesthesiaAnesthesia & Analgesia, 1995
- Adverse Respiratory Events in Anesthesia: A Closed Claims AnalysisAnesthesiology, 1990