The imaging performance of a storage phosphor system for dental radiography
- 1 March 1996
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 69 (819) , 256-261
- https://doi.org/10.1259/0007-1285-69-819-256
Abstract
Direct digital imaging systems are becoming increasingly common in dental radiography. The majority of these systems are based on charge coupled device (CCD) technology. A relatively new system, based on photo-stimulable phosphor luminescence (PSPL), is now commercially available. This system appears to overcome some of the restrictions of CCD systems including those associated with the bulky detector, connecting wire, limited image size and limited exposure latitude. A physical evaluation of the PSPL system was conducted including measurements of sensitometric response, modulation transfer function (MTF), noise power spectrum (NPS), noise equivalent quanta (NEQ) and detective quantum efficiency (DQE). These measurements were compared with results obtained previously for Kodak Ektaspeed (E-speed) film. The results of the evaluation indicate that the PSPL system may be capable of operating at exposures up to 80% lower than for film with comparable image quality.Keywords
This publication has 4 references indexed in Scilit:
- Signal, noise and SNR transfer properties of computed radiographyPhysics in Medicine & Biology, 1993
- Physical aspects of photostimulable phosphor computed radiographyThe British Journal of Radiology, 1993
- Quantitative assessment of a new dental imaging systemThe British Journal of Radiology, 1991
- The Specification of Imaging Properties by Response to a Sine Wave InputJournal of the Optical Society of America, 1954