Abstract
An investigation of the signal, noise and signal-to-noise transfer properties of a Fuji 7000 series computed radiography (CR) system using third-generation storage plate technology has been undertaken. The results show a significant improvement in signal-to-noise performance of the system compared to an earlier generation of CR technology. Results of the CR evaluation compare favourably to those obtained from a selection of rare-earth screen-film systems evaluated under identical exposure conditions. These results were further substantiated by results from threshold contrast-detail detectability tests.Keywords
This publication has 22 references indexed in Scilit:
- Threshold perception performance with computed and screen-film radiography: implications for chest radiography.Radiology, 1992
- A physical image quality evaluation of a digital spot fluorography systemPhysics in Medicine & Biology, 1992
- Digital X-ray imagingMeasurement Science and Technology, 1991
- Quantitative radiographic imaging using a photostimulable phosphor systemMedical Physics, 1990
- Resolution Properties Of A Computed Radiographic SystemPublished by SPIE-Intl Soc Optical Eng ,1989
- Basic imaging properties of a computed radiographic system with photostimulable phosphorsMedical Physics, 1989
- Low Print-Through Technology With Rare Earth Tantalate PhosphorsPublished by SPIE-Intl Soc Optical Eng ,1985
- Analysis of variations in contrast‐detail experimentsMedical Physics, 1984
- Quantitative relation between detectability and noise powerPhysics in Medicine & Biology, 1983
- Detection Of Bars And Discs In Quantum NoisePublished by SPIE-Intl Soc Optical Eng ,1979