Dosimetry of digital panoramic imaging. Part I: patient exposure
- 1 May 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Dento maxillo facial radiology
- Vol. 34 (3) , 145-149
- https://doi.org/10.1259/dmfr/28107460
Abstract
To measure patient radiation dose during panoramic exposure with various panoramic units for digital panoramic imaging. An anthropomorphic phantom was filled with thermoluminescent dosemeters (TLD 100) and exposed with five different digital panoramic units during ten consecutive exposures. Four machines were equipped with a direct digital CCD (charge coupled device) system, whereas one of the units used storage phosphor plates (indirect digital technique). The exposure settings recommended by the different manufacturers for the particular image and patient size were used: tube potential settings ranged between 64 kV and 74 kV, exposure times between 8.2 s and 19.0 s, at fuse current values between 4 mA and 7 mA. The effective radiation dose was calculated with inclusion of the salivary glands. Effective radiation doses ranged between 4.7 microSv and 14.9 microSv for one exposure. Salivary glands absorbed the most radiation for all panoramic units. When indirect and direct digital panoramic systems were compared, the effective dose of the indirect digital unit (8.1 microSv) could be found within the range of the effective doses for the direct digital units (4.7-14.9 microSv). A rather wide range of patient radiation doses can be found for digital panoramic units. There is a tendency for lower effective doses for digital compared with analogue panoramic units, reported in previous studies.Keywords
This publication has 22 references indexed in Scilit:
- Quality of film-based and digital panoramic radiographyDento maxillo facial radiology, 2004
- Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unitDento maxillo facial radiology, 2003
- Accuracy of digital and film panoramic radiographs for assessment of position and morphology of mandibular third molars and prevalence of dental anomalies and pathologiesDento maxillo facial radiology, 2003
- An introduction to digital radiography in dentistryJournal of Orthodontics, 2002
- Radiation dose in dental radiologyEuropean Radiology, 2002
- Digital Imaging for DentistsAustralian Dental Journal, 2001
- Effective dose from radiation absorbed during a panoramic examination with a new generation machineOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2000
- An update on dental imagingBritish Dental Journal, 1998
- Charge-coupled device panoramic radiography: effect of beam energy on radiation exposure.Dento maxillo facial radiology, 1998
- 1992 assessment of radiation risk from dental radiography.Dento maxillo facial radiology, 1992