Morphological changes and lens position 2½ years after intracapsular cataract extraction with implantation of a semiflexible anterior chamber lens
- 1 April 1987
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 65 (2) , 237-243
- https://doi.org/10.1111/j.1755-3768.1987.tb07006.x
Abstract
Morphological changes and lens position were examined in 51 patients 25 to 38 (mean 30) months after intracapsular cataract extraction (ICCE) with implantation of the semiflexible, 3M, style 70, anterior chamber lens. Morphological changes presumably associated with the ICCE proper were: iris transillumination defects (68.6%), rupture of anterior hyaloid membrane (20%), detachment of posterior vitreous (42.9%) and anterior synecchiae (13.7%). The implant-related morphological changes were caused by iris tucking (19.6%), small, mobile lenses causing endothelial disturbance (9.8%), iris-haptic adhesions and progressive ovalling of pupullary form. The present morphological changes were compared with a previously published examination of the same patients performed 4 months after surgery: 9 eyes without tissue changes around the lens feet at the previous follow-up now had iris-haptic adhesions involving from 1 to 4 lens feet. In 8 eyes the iris-haptic adhesions had progressed. Compared with the previous examination the fraction of eyes with tissue changes at the lens feet had risen from 47.1 to 64.7% while the number of eyes with oval pupils had increased from 54.9 to 80.4%. Contact between lens haptic and iris root seems to constitute a constant stimulus for structural changes of the iris.Keywords
This publication has 16 references indexed in Scilit:
- Irreversible Intraocular Lens Vaulting with Pupillary-Block GlaucomaAmerican Journal of Ophthalmology, 1986
- Pupillary elongation after anterior chamber lens implantationJournal of Cataract & Refractive Surgery, 1986
- Intraocular pressure in the first days after implantation of the semiflexible McGhan/3 M, style 70, anterior chamber lensActa Ophthalmologica, 1985
- Pathological and scanning electron microscopic evaluation of the 91Z intraocular lensAmerican Intra-Ocular Implant Society Journal, 1984
- Evaluation of semiflexible and flexible anterior chamber intraocular lensesAmerican Intra-Ocular Implant Society Journal, 1983
- Vaulting Characteristics of Flexible Loop Anterior Chamber Intraocular LensesArchives of Ophthalmology (1950), 1983
- Dislocation of the Kelman II anterior chamber intraocular lensAmerican Intra-Ocular Implant Society Journal, 1982
- A study of the anterior chamber before and after implantation of the Choyce Mark VIII lensAmerican Intra-Ocular Implant Society Journal, 1980
- A TECHNICAL IMPROVEMENT OF THE HAAG‐STREIT PACHOMETER Short CommunicationActa Ophthalmologica, 1977
- Vitreous Changes Produced by Cataract SurgeryArchives of Ophthalmology (1950), 1966