Effects of Time to Defibrillation and Subthreshold Preshocks on Defibrillation Success in Pigs
- 1 February 1989
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 12 (2) , 358-365
- https://doi.org/10.1111/j.1540-8159.1989.tb02669.x
Abstract
The purpose of this study was to examine the effects of: (1) time to defibrillation and (2) subthreshold preshocks on defibrillation success. We conducted two separate experiments in 19 anaesthetized, open-chested pigs. Defibrillation was attempted using the sequential pulse technique approximately 10 s after electrically induced ventricular fibrillation. Each sequential pulse shock consisted of two trapezoidal pulses (approximately 3 ms duration), separated by 0.2 ms. Current was delivered to three mesh electrodes (TX-7, Medtronic) sutured over the anterior right ventricle, posterior right ventricle, and lateral left ventricle. For each animal, defibrillation threshold (DFT) defined as the lowest delivered energy that defibrillated the heart, was measured twice and the average was designated as mean DFT. The energy at 1.4 times the specific mean DFT for each pig was designated as the test shock and 100 volts less than the mean DFT was designated as the preshock. In the first experiment, test shocks were delivered at five different fibrillation intervals (10, 20, 40, 60, and 90 s). Time to test shock delivery was randomized for 10 to 60 s, but, 90 s was always tested at the end of the study. Percentages of success at 10, 20, 40, 60, and 90 s were 94, 78, 94, 83, and 100%, respectively (p = NS). The 12 pigs in which all initial test shocks were successful were selected for the evaluation of post defibrillation arrhythmias. The cumulative incidence of complete heart block lasting at least 5 s were 8, 33, 83, 83, and 83%, respectively. The incidence of complete heart block increased significantly at 40 s (p less than 0.001).(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 6 references indexed in Scilit:
- Defibrillation Threshold: A Simple and Quantitative Estimate of the Ability to DefibrillatePacing and Clinical Electrophysiology, 1987
- Defibrillation with the sequential pulse technique: Reproducibility with repeated shocksAmerican Heart Journal, 1986
- Internal cardiac defibrillation in man: pronounced improvement with sequential pulse delivery to two different lead orientations.Circulation, 1986
- The automatic implantable cardioverter-defibrillator: An overviewJournal of the American College of Cardiology, 1985
- Improved internal defibrillation with twin pulse sequential energy delivery to different lead orientations in pigsThe American Journal of Cardiology, 1985
- Ventricular Defibrillation — A Comparative Trial Using 175-J and 320-J ShocksNew England Journal of Medicine, 1982