Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomography
- 14 September 2006
- journal article
- research article
- Published by Wiley in Clinical & Experimental Ophthalmology
- Vol. 34 (8) , 751-754
- https://doi.org/10.1111/j.1442-9071.2006.01343.x
Abstract
Measurement of central corneal thickness (CCT) plays an important role in both diagnostic and therapeutic assessment of ocular diseases. Although ultrasound pachymetry (U-PACH) is regarded as the golden standard for measurement of CCT, optical coherence tomography (OCT) may offer advantages as it can locate the central cornea with precision with no corneal touch. Nevertheless, the agreement of OCT with U-PACH has not yet been gauged by Bland-Altman analysis. This study compares CCT measurement by OCT with that by U-PACH. Healthy subjects without ocular abnormality (except refractive errors less than or equal to -6.0 D), contact lens wear or ocular surgery were recruited. CCT was measured in one eye of normal subjects using OCT and U-PACH. Results were compared using correlation and Bland-Altman plots. Fifty subjects were recruited. Mean +/- SD CCT measured by OCT was 565 +/- 33 microm. This was highly correlated (Pearson's coefficient = 0.934) with the mean thickness measured by U-PACH (543 +/- 33 microm). The coefficients of variation were good and comparable at 7.9% for U-PACH and 3.5% for OCT. Compared with U-PACH, OCT consistently overestimated the CCT by a mean of 23 microm as shown on Bland-Altman plot. CCT measured by OCT and U-PACH is highly correlated. With appropriate adjustment factor, OCT agrees well with U-PACH and is a reliable alternative for CCT measurement.Keywords
This publication has 16 references indexed in Scilit:
- Central corneal pachymetry and visual field progression in patients with open-angle glaucoma☆Ophthalmology, 2004
- Clinical Significance of Central Corneal Thickness in the Managementof GlaucomaArchives of Ophthalmology (1950), 2004
- Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetryEye, 2002
- The Repeatability of Corneal Thickness MeasuresCornea, 2000
- Optical coherence tomography evaluation of the corneal cap and stromal bed features after laser in situ keratomileusis for high myopia and astigmatismOphthalmology, 2000
- Corneal thickness and curvature in normal-tension glaucomaAmerican Journal of Ophthalmology, 1998
- Statistics Notes: Measurement errorBMJ, 1996
- Confocal Microscopy of the EyePublished by Springer Nature ,1990
- Differences Among Ultrasonic Pachymeters in Measuring Corneal ThicknessJournal of Refractive Surgery, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986