Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro models
- 1 November 2001
- journal article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 19 (6) , 1162-1167
- https://doi.org/10.1016/s0736-0266(01)00039-0
Abstract
Radiostereometric analysis (RSA) is used to measure translations of joint replacement components with respect to the host bone in vivo. We used two cadaveric models of hip arthroplasty, one human and one canine, to evaluate the accuracy and precision of RSA-based estimates of translations of the femoral component with respect to the femur under ideal conditions. The femoral components were attached rigidly to a micrometer stage that provided standard displacements in increments of 25 and 50 microm in the interval from zero to 500 microm along three orthogonal axes. Radiostereometric examinations were performed for each increment. Accuracy was calculated as the 95% prediction intervals from regression analyses between the measured and actual displacements. Precision was evaluated as the standard deviation of five repeated measurements of a 200 microm displacement along each axis. Both accuracy and precision were best along the longitudinal axis, with a prediction interval of +/-47 microm in the human model and +/-45 microm in the canine model and a standard deviation of 30 microm in the human model and 40 microm in the canine model. The use of only the prosthetic head as a landmark (as opposed to three markers placed on the femoral stem) led to a 3-fold larger prediction interval in the human model and a 2-fold greater prediction interval in the canine model.Keywords
This publication has 11 references indexed in Scilit:
- Determining the movements of the skeleton using well-configured markersPublished by Elsevier ,2004
- Error propagation for relative motion determined from marker positionsJournal of Biomechanics, 1997
- Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prosthesesThe Journal of Bone and Joint Surgery. British volume, 1995
- Micromotion of femoral stems in total hip arthroplasty. A randomized study of cemented, hydroxyapatite-coated, and porous-coated stems with roentgen stereophotogrammetric analysis.Journal of Bone and Joint Surgery, 1994
- Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prosthesesThe Journal of Bone and Joint Surgery. British volume, 1994
- Migration of acetabular components, inserted with and without cement, in one-stage bilateral hip arthroplasty. A controlled, randomized study using roentgenstereophotogrammetric analysis.Journal of Bone and Joint Surgery, 1994
- Early loosening of revision hip arthroplasty: A roentgen stereophotogrammetric analysisThe Journal of Arthroplasty, 1990
- Roentgen stereophotogrammetryActa Orthopaedica, 1989
- Mechanical loosening of total hip prostheses. A radiographic and roentgen stereophotogrammetric studyThe Journal of Bone and Joint Surgery. British volume, 1986
- Micromotion in knee arthroplastyActa Orthopaedica, 1986