Longitudinal precision of dual-energy X-ray absorptiometry in a multicenter study
- 1 February 1991
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 6 (2) , 191-197
- https://doi.org/10.1002/jbmr.5650060213
Abstract
Reproducibility is a key issue in both clinical and research applications of bone mineral density (BMD) measurements. To examine the longitudinal precision of dual‐energy x‐ray absorptiometry (DEXA) for the measurement of mineral density in vivo and in vitro, the performance of a group of instruments in the course of a multicenter longitudinal clinical trial was monitored. Measures were performed on eight identical machines (Hologic QDR1000) and analyzed using the same automated software program. Short‐term precision was good in vitro, [anthropomorphic spine phantoms; mean intrasite coefficient of variation (CV) 0.42 ± 0.1% (SD)] and in vivo (lumbar spine; CV 1.1 ± 0.5%). Intersite measures of a single spine phantom (specified mineral content −57.8 g) revealed a range of 57.3–58.4 g (CV 0.7%). In two subjects intersite CV in vivo were 3.7 and 2.1% (spine) and 1.8 and 3.2% (femoral neck). At five sites frequent phantom measures were performed over a 1 year period (mean number of measures 196) and revealed a mean all‐point CV of 0.43% (range 0.35–0.53%). Longitudinal precision in vivo was somewhat less (mean CV of spinal measures 1.1%, femoral neck 1.2%, trochanter 1.3%, and Ward's area 2.4%). At one additional site large variations in phantom measures heralded repeated mechanical failures that eventually required machine replacement. In summary, DEXA demonstrates good in vitro and in vivo longitudinal precision, providing the basis for expanded clinical and research usefulness. Nevertheless, stringent quality assurance measures are required to detect and respond to system malfunctions.Keywords
Funding Information
- Syntex Laboratories and Veterans Administration Merit Review Funds
- Submitted to American Society for Bone
- Mineral Research Twelfth Annual Meeting, Atlanta, Georgia
This publication has 16 references indexed in Scilit:
- In vivo long-term precision of spinal bone mass measurement by dual photon absorptiometry Dear editorBone and Mineral, 1989
- Performance evaluation of a dual-energy X-ray bone densitometerCalcified Tissue International, 1989
- Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar SpineMayo Clinic Proceedings, 1988
- New Radiographic Technique May Renew Credibility of Bone DensitometryMayo Clinic Proceedings, 1988
- Quantitative Digital RadiographyVersusDual Photon Absorptiometry of the Lumbar Spine*Journal of Clinical Endocrinology & Metabolism, 1988
- The clinical relevance of calcaneus bone mineral measurements: a reviewBone and Mineral, 1988
- Dual energy radiography (DER): A preliminary comparative studyCalcified Tissue International, 1988
- Measurement of spinal or peripheral bone mass to estimate early postmenopausal bone loss?The American Journal of Medicine, 1988
- Bone mass measurement: Which site to measure?The American Journal of Medicine, 1988
- Quantitative CT for determination of bone mineral density: a review.Radiology, 1988