Limitations of the point and line source approximations for the determination of geometry factors around brachytherapy sources
- 6 January 2000
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 27 (1) , 124-128
- https://doi.org/10.1118/1.598873
Abstract
Geometry factors were calculated around commercially available pellets and elongated brachytherapy sources taking into account their actual active core geometries. These calculations were compared with corresponding ones derived using the point and line source approximations commonly used for the determination of geometry factors, as proposed by AAPM Task Group 43. The point source approximation was found to be efficient for the determination of geometry factors around single active pellets, even at radial distances very close to the source. It is also valid for the determination of geometry factors at radial distances from elongated brachytherapy sources of length For smaller radial distances, however, this approximation introduces significant errors, >50%, around elongated source designs, thus being unacceptable for the determination of geometry factors in this case. The line source approximation was found to accurately reproduce geometry factors around elongated brachytherapy source designs. Errors greater than 3%, due to the fact that the line source approximation ignores the radial dimension d of the source, are observed only at radial distances very close to the source at polar angles far away from their transverse bisectors. These errors depend on the ratio and increase as this ratio increases.Keywords
This publication has 12 references indexed in Scilit:
- A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy sourceMedical Physics, 1999
- Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60Medical Physics, 1999
- Monte Carlo‐aided dosimetry of a new high dose‐rate brachytherapy sourceMedical Physics, 1998
- Monte Carlo and TLD dosimetry of an high dose‐rate brachytherapy sourceMedical Physics, 1998
- Application of the Monte Carlo integration (MCI) method for calculation of the anisotropy of brachytherapy sourcesPhysics in Medicine & Biology, 1998
- Monte Carlo dosimetry of the VariSource high dose rate 192Ir sourceMedical Physics, 1998
- Monte Carlo calculation of dose rate distributions around wiresMedical Physics, 1997
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Comparison of measured and calculated dose rates in water near I‐125 and Ir‐192 seedsMedical Physics, 1991
- Dose parameters of 125I and 192Ir seed sourcesMedical Physics, 1989