AEC set-up optimisation with computed radiography imaging
- 1 December 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 117 (1-3) , 169-173
- https://doi.org/10.1093/rpd/nci743
Abstract
Phototimer set-up is a critical procedure for dose and image quality optimisation in computed radiography (CR) systems. While a conventional radiography automatic exposure control device (AEC) can be calibrated in order to gain a constant optical density on the film independent of beam quality and patient size, CR detectors present a high dynamic range which allows a much larger dose interval, but with different image quality levels. CR leads to a less frequent exam repetition, but may produce quite noisy images if the exposure level on the plate is not correct. The aim of this work is to evaluate the performance of a CR plate (Agfa MD40) in order to optimally calibrate an AEC device. The plate response has been characterised in terms of digital signal, exposure on the plate and signal-to-noise ratio for different beam qualities, in a patient of standard size.Keywords
This publication has 8 references indexed in Scilit:
- Phototimer setup for CR imagingMedical Physics, 2000
- Comparison of Doses for Bedside Examinations of the Chest with Conventional Screen-Film and Computed Radiography: Results of a Randomized Controlled TrialRadiology, 2000
- Relative speeds of Kodak computed radiography phosphors and screen-film systemsMedical Physics, 1997
- Computed radiography X-ray exposure trendsAcademic Radiology, 1996
- Mottle on computed radiographs of the chest in pediatric patients.Radiology, 1996
- Threshold perception performance with computed and screen-film radiography: implications for chest radiography.Radiology, 1992
- Comparison of imaging properties of a computed radiography system and screen–film systemsMedical Physics, 1991
- SNR and DQE analysis of broad spectrum X-ray imagingPhysics in Medicine & Biology, 1985