A novel method for analysis of TOMROP data
Open Access
- 29 March 2004
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 19 (4) , 508-512
- https://doi.org/10.1002/jmri.20027
Abstract
Purpose To develop an efficient method for extracting maps of the corrected T1 from images generated using the T One by Multiple Read Out Pulses (TOMROP) sequence. Materials and Methods An expression is developed for the true T1 in terms of the parameters from a three-parameter fit of the TOMROP data. Solutions of gadodiamide in normal saline with concentrations of 0.0, 0.06, 0.11, 0.23, 0.46, and 0.91 mM were prepared and T1 measurements were performed using both the inversion recovery (IR) and the TOMROP methods. The TOMROP data were analyzed using the proposed technique and the results compared to those from the IR measurements. Results The T1 estimates generated from the TOMROP data using the proposed method were consistent with the IR results. However, systematic errors were observed in the T1 estimates when the repetition time was not sufficient for full recovery of axial magnetization. Relaxation times determined using the proposed method were within 1% of the spectroscopiclly determined values for T1 values in the range of 0.28–2.8 seconds when a suitable delay was employed. Conclusion The proposed method of analysis was found to yield accurate T1 estimates when the assumptions used in the analysis were not violated. J. Magn. Reson. Imaging 2004;19:508–512.Keywords
This publication has 10 references indexed in Scilit:
- Relaxographic ImagingJournal of Magnetic Resonance, Series B, 1994
- Accurate measurement of T1in vivo in less than 3 seconds using echo‐planar imagingMagnetic Resonance in Medicine, 1993
- Fast and accurate measurements of T1 using a multi‐readout single inversion‐recovery sequenceMagnetic Resonance in Medicine, 1992
- Quantification of T1 values by SNAPSHOT-FLASH NMR imagingJournal of Magnetic Resonance (1969), 1992
- Inversion Recovery Snapshot FLASH MR ImagingJournal of Computer Assisted Tomography, 1989
- A comparison of one‐shot and recovery methods in T1 imagingMagnetic Resonance in Medicine, 1988
- Strategies and tactics in NMR imaging relaxation time measurements. I. Minimizing relaxation time errors due to image noise—the ideal caseMagnetic Resonance in Medicine, 1985
- The choice of optimal parameters for measurement of spin-lattice relaxation times. IV. Effects of nonideal pulsesJournal of Magnetic Resonance (1969), 1985
- NMR population inversion using a composite pulseJournal of Magnetic Resonance (1969), 1979
- Time Saving in Measurement of NMR and EPR Relaxation TimesReview of Scientific Instruments, 1970