Accurate measurement of T1in vivo in less than 3 seconds using echo‐planar imaging
- 1 September 1993
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 30 (3) , 351-354
- https://doi.org/10.1002/mrm.1910300312
Abstract
This paper describes a rapid and accurate method for meassuring T1 in vivo using an echo-planar imaging version of the Look-Locher sequence. T1 values from 76 to 1330 ms have been measured in 3 s with a mean accuracy of 4.6%.Keywords
This publication has 12 references indexed in Scilit:
- Fast and accurate measurements of T1 using a multi‐readout single inversion‐recovery sequenceMagnetic Resonance in Medicine, 1992
- Experimental analysis of T1 imaging with a single‐scan, multiple‐point, inversion‐recovery techniqueMagnetic Resonance in Medicine, 1992
- Practical Implementation and Optimization of One‐shot T1 imagingMagnetic Resonance in Medicine, 1991
- A simple method for the restoration of signal polarity in multi‐image inversion recovery sequences for measuring T1Magnetic Resonance in Medicine, 1991
- High‐speed multislice T1 mapping using inversion‐recovery echo‐planar imagingMagnetic Resonance in Medicine, 1990
- Snapshot imaging at 0.5 t using echo‐planar techniquesMagnetic Resonance in Medicine, 1989
- A comparison of one‐shot and recovery methods in T1 imagingMagnetic Resonance in Medicine, 1988
- Measurement of T1by echo-planar imaging and the construction of computer-generated imagesPhysics in Medicine & Biology, 1986
- Multi-planar image formation using NMR spin echoesJournal of Physics C: Solid State Physics, 1977
- Time Saving in Measurement of NMR and EPR Relaxation TimesReview of Scientific Instruments, 1970