Off-axis beam quality change in linear accelerator x-ray beams
- 1 March 1980
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 7 (2) , 145-146
- https://doi.org/10.1118/1.594677
Abstract
The effective energy of the X-ray beam from radiotherapy linear accelerators changes as a function of the position in the beam due to nonuniform filtration by the flattening filter. The transmittance through a water column was measured in good geometry and the beam quality characterized in units of HVL [half-value layer] in water. Measurements were made on a variety of linear accelerators from 4-10 MV. The beam energy decreased with increasing distance from the central ray for all accelerators measured.Funding Information
- National Cancer Institute (CA 10953)
- DHEW
This publication has 1 reference indexed in Scilit: